beautypg.com

Annually, Sheath filter, As required – Luminex 200 System User Manual User Manual

Page 42: Fuses

background image

Annually

Sheath Filter

1. Disconnect the Luminex 200 analyzer from AC power by turning off the power switch on

the rear of the analyzer, then unplugging the analyzer power cord from the wall source.

2. Disconnect the sheath fluid tubing before changing the filter.

3. Open the left door on the Luminex 200 analyzer. Disconnect the filter by pushing down

on the metal clamps on each connection. See the following figure.

FIGURE 11.

Sheath Filter

4. Connect the new sheath filter, matching up the color-coded fittings. The arrow on the

sheath filter should be pointing up.

5. Reconnect the sheath fluid tubing.

6. Plug in and power on the Luminex 200 analyzer.

7. Close the left analyzer door.

8. Prime twice.

As required

Fuses

WARNING:

To avoid serious injury or death by electric shock, you must turn
off the system and unplug it from the wall.

The following procedure applies to both the Luminex 200 analyzer and the Luminex XYP
instrument.

Luminex

®

200

System User Manual

30

For In Vitro Diagnostic Use