Sensitivity, Capacity, Luminex 200 analyzer general – Luminex 200 System User Manual User Manual

Page 24: Sensitivity capacity

background image

• Internal sample carry over: < 0.9%

• Soluble background fluorescence emission at 575 nm automatically subtracted from

fluorescence intensity values

Sensitivity

• Detect 1000 fluorochromes phycoerythrin (PE) per xMAP microsphere

• Reporter channel dynamic range: 3.5 decades of detection

Capacity

The specifications below reflect minimum capacity values:

• Analyze multiple 96-well plates per batch

• Analyze multiple assay templates per plate

• Distinguish a minimum of 1 to a maximum of 100 unique xMAP microsphere sets in a

single sample

• Detect and distinguish surface reporter fluorescence emissions at 575 nm on the surface of

1-100 unique xMAP microspheres sets in a single sample

• Sample core: 15-20 µm core at 1 µL/sec. sample inject rate

• Maintain samples at a constant temperature from 35°C to 55°C (95°F to 131°F)

• Automatic sampling from a 96-well plate

• Start sampling from any well position

• Sheath container and waste container hold enough volume to run up to two 96-well plates

between refills

• Microtiter plates with 96 wells must be compatible with the Luminex XYP instrument plate

holder. The following microtiter plate types are compatible with the Luminex XYP
instrument plate holder: flatbottom, conical, round, filter bottom, half plates, overall height
no more than 0.75” (19 mm), any color.

• Microtiter plates with 96 wells must be compatible with Luminex XYP instrument heater

block temperature from 35°C to 55°C (95°F to 131°F) when performing heated assays and
using the heater block.

Luminex 200 Analyzer General

• Indoor use only

• Operating temperature: 15°C to 30°C (59°F to 86°F)

• Humidity: 20% to 80%, noncondensing

• Altitude: Operation up to 2400 m (7874 ft.) above mean sea level

• Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches) D x 24.5 cm (9.5 inches)

H

• Weight: maximum of 25 kg (60 lbs.)

• UL installation category: UL Installation Category II, as defined in Annex J of UL 61010A-1

• Pollution degree: UL Pollution Degree 2, as defined in Section 3.7.3.2 of UL 61010A-1

Luminex

®

200

System User Manual

12