background image

Sheath and Waste Fluids

Replace the sheath fluid and empty the waste container as required. Be careful not to touch
the seal in the waste bottle cap or allow it to become wet or dirty. This could cause the bottle
to become pressurized, causing pressure errors in the system. If the seal gets wet, let it air-
dry. Touching the seal can contaminate it. If it gets dirty, you should replace the waste bottle
cap.

If you are using the waste line, do not move it while the system is running. While you can
move the line along a horizontal surface, do not permanently reroute the waste line elevation
without first contacting Technical Support. You can move the line temporarily for cleaning and
maintenance purposes.

You must manually monitor waste container levels.

Refill the Sheath Fluid Container

To refill the sheath fluid container:

1. Release system pressure by removing the lid from the sheath fluid container.

2. Refill the sheath fluid container.

If the sheath container ever runs dry, prime the system at least twice until the air is
removed from the system.

Empty the Waste Container

To empty the waste container:

1. Disconnect the waste container from the Luminex 200 analyzer.

2. Unscrew the waste container lid, being careful not to touch the Gore-Tex

seal. If the

seal becomes wet or soiled, it may impair venting.

3. Discard the waste from the waste container by appropriate means.

4. Reconnect the waste container to the Luminex 200 analyzer and replace the cap.

NOTE: There is NO warning of a high waste volume. Empty the waste

container each time you fill the sheath container.

Any time the sheath container is disconnected from the Luminex 200 analyzer, you must
remove air from the sample lines by priming.

Weekly

Visual Inspection

Open all of the Luminex 200 analyzer doors and visually inspect for leaks, corrosion, and
other signs of improper function. Check all visible tubing connections. Check the Luminex
XYP instrument air intake filter for buildup of dust. Check the SD system and its connection
for leaks. If you see a leak, turn off the power to the Luminex SD system and contact
Luminex Corporation.

For In Vitro Diagnostic Use

Maintenance and Cleaning