Luminex XYP Instrument Air Intake Filter

1. Disconnect the Luminex XYP instrument from AC power by turning off the power switch

on the rear of the Luminex XYP instrument, then unplugging the Luminex XYP instrument
power cord from the wall source.

2. On the back of the Luminex XYP instrument, to the left side, gently remove the screen

from the Luminex XYP instrument filter.

NOTE: Do not remove the screws.

FIGURE 7.

Removing the Screen

3. Replace the filter and replace the screen.

FIGURE 8.

Removing and Replacing the Filter

4. Plug in and turn on Luminex XYP instrument power.

For In Vitro Diagnostic Use

Maintenance and Cleaning