beautypg.com

Luminex sd system general, Pc specifications, Recommended additional equipment – Luminex 200 System User Manual User Manual

Page 26: Uninterruptible power supply (ups), Surge protector, Printer, Barcode labels

background image

Luminex SD System General

• Ambient temperature: 15°C to 30°C (59° to 86°F)

• Humidity: 20% to 80%, noncondensing

• Altitude: designed to operate at up to 2400m (7874 feet) above mean sea level

• Physical dimensions: 20 cm (8 inches) W x 30 cm (11.75 inches) D x 24.75 cm (9.75

inches) H

• Weight: 9 kg (20 lbs)

• UL installation category: UL Installation Category II, as defined in Annex J of UL 61010A-1

• Pollution degree: UL Pollution Degree 2, as defined in Section 3.7.3.2 of UL 61010A-1

• Input voltage range: 100-240 V~ ± 10%, 0.4 Amps, 47-63 Hz

• AC inlet fuse: 2 Amp, 250 V~, time lag

PC Specifications

For updated information regarding the PC, notebook, or operating system, go to http://
www.luminexcorp.com, then click on the Support link to open the FAQ list.

Recommended Additional Equipment

Uninterruptible Power Supply (UPS)

Luminex highly recommends using an uninterruptible power supply (UPS) to protect your
system from power outages. Choose one that can provide 1050 Watts for at least 45 minutes.
The UPS should be UL listed (or equivalent), CSA certified, and CE marked when used
internationally.

Surge Protector

If you do not use a UPS, use a surge protector. Choose a protector that meets your needs.
Factors to consider include electrical environment, endurance, suppressed voltage rating,
and method of protection. It should have six outlets, rated at least 1500 Watts, and be UL
listed, CSA certified, CE marked for nondomestic use when used internationally.

Printer

Printer, HP LaserJet 2300 or available equivalent

Barcode Labels

Use Code 128 barcode label type when scanning barcode labels into the system.

Luminex

®

200

System User Manual

14

For In Vitro Diagnostic Use