Measurement device technical data
Subject to changes.
Performance specifications
Measurement method
Reduction method, derived from DIN 38405 D11,
analysis photometric via IR LED photometer
Measurement ranges
0.01–0.50...5.00 mg/L P total phosphorus in 0.1 mg/L increments
0.01–0.50...5.00 mg/L P orthophosphate in 0.1 mg/L increments
(depending on design)
Measurement uncertainty
0.01 – 5.00 mg/L: ±2 % from the measurement value
Measurement interval
t
100
= 10 min
Sample requirement
approximately 100 mL/h
Ambient temperature
+ 5° C to + 40° C
Calibration
Automatic at selectable intervals
Inspection interval
3 months
Maintenance requirement
approximately 1/2 hour per week, typical
Reagent stock
3 months (standard solution 6–12 months)
Outputs
2 current outputs:
2 limit-value contacts:
Service interface:
Interface:
0/4...20 mA, maximum 500 Ω
potential-free, 24 V 1A
RS 232
bus-capable (optional)
Protection type
in accordance with IP 54
Mains connection
230 V AC / 50 Hz
Power consumption
Approximately 150 VA
Dimensions: W x H x D
560 mm x 810 mm x 330 mm
Mass
Approximately 29 kg (without reagents)
Fuses
T2.5 A L; 250 V (2 units)
Other
Automated cleaning, data logger, graphic display with load curve display