Fluid Components International CMF Series Manual Electronics Setup and Menus User Manual
Page 8
Commissioning
FlexCOR
™
Model CMF Series
58
Fluid Components Intl
Document 06EN003327 Rev. -
Menu detail
Digital output
Limit/direction
Limit switches are available for both digital and relay output and can be used for mass flow, fraction,
volume flow, temperature or density.
To set the output to show flow direction then select: 1 set point at 0% flow; hysteresis 5%.
Digital output
Error level
Acceptance level is set in the basic settings.
Digital output
Error number
Digital output
Flow direction
1 set point
To have a flow, temperature, density or fraction area monitored via the digital output then select
2 set points.
2 set points
If 2 set points have to activate two separate outputs, each set point has to be selected individually
one for digital output and the other for relay output.
2 separate set points
If the error statusis to be output at the digital output there
are two options. Select error level to output the system’s
current error status or error number if a specific error is
to be indicated by the digital output.
Both types of error message are described in more
detail in the section error system.