Commissioning

FlexCOR

Model CMF Series

58

Fluid Components Intl

Document 06EN003327 Rev. -

Menu detail

Digital output

Limit/direction

Limit switches are available for both digital and relay output and can be used for mass flow, fraction,

volume flow, temperature or density.

To set the output to show flow direction then select: 1 set point at 0% flow; hysteresis 5%.

Digital output
Error level

Acceptance level is set in the basic settings.

Digital output

Error number

Digital output

Flow direction

1 set point

To have a flow, temperature, density or fraction area monitored via the digital output then select
2 set points.

2 set points

If 2 set points have to activate two separate outputs, each set point has to be selected individually
one for digital output and the other for relay output.

2 separate set points

If the error statusis to be output at the digital output there

are two options. Select error level to output the system’s

current error status or error number if a specific error is
to be indicated by the digital output.

Both types of error message are described in more

detail in the section error system.