Commissioning

Fluid Components Intl

Document 06EN003327 Rev. -

69

FlexCOR

Model CMF Series

7.6.3 Batch

A batch can be started, paused or stopped from the operator menu, in addition to the externally

operated batch control. The batch is controlled using the lock and the top up keys.

The lock key:

-

Starts the batch

-

Holds the batch (pause) when pressed during batching

-

Continue the batch when pressed during a pause.

The top up key resets a batch completely during a pause.

The accumulated number of performed batches can be viewed when enabled in the operator
menu setup.

The batch cycle counter is reset by pressing the lock key in the “batch cycle cnt reset” menu.

Batch cycle counter

Batch cycle counter reset

7.6.1 Operator menu

Flowrate

The 1

st

displayline will always be active and show the value enabled in the operator menu setup.

-

Massflow rate, volumeflow rate, density, temperature, totalizer1, totalizer2

-

Totalizer 2

The 2

nd

and 3

th

display lines are individually set in the operator menu. The page forward key steps

through the enabled settings.

-

Mass flow rate

-

Volume flow rate

-

Density

-

Temperature

-

Totalizer

-

Totalizer reset

-

Batch control

-

Batch cycle counter

-

Batch cycle counter reset.

7.6.2 Totalizer

A totalizer is reset by pressing the lock key when the corresponding totalizer reset window is open.