Commissioning

FlexCOR

Model CMF Series

70

Fluid Components Intl

Document 06EN003327 Rev. -

Parameter

Factory setting

Settings available

Password

Password

1000

1000 – 9999

Basic settings

Flow direction

Positive

Positive, negative

Massflow max.

Dim. dependent

Dim. dependent

- Weight units

Dim. dependent

mg, g, kg, ton, lb

- Time units

Dim. dependent

s, min, h, d

Volumeflow max.

Dim. dependent

Dim. dependent

- Volume units

Dim. dependent

m

3

, ml, l, hl, kl, Ml, ft

3

, in

3

, US G, US MG,

UK G, UK MG

- Time units

Dim. dependent

s, min, h, d

Density

- Minimum

+0.1 g/cm

3

-20000.0 kg/m

3

- +20000.0 kg/m

3

- Maximum

+2.0 g/cm

3

-20000.0 kg/m

3

- +20000.0 kg/m

3

- Weight units

g

mg, g, kg, ton, lb

- Time units

cm

3

cm

3

, m

3

, ft

3

, in

3

Sensor temp.

- Minimum

-50 °C

-250 °C - +250 °C

- Maximum

+250 °C

-250 °C - +250 °C

- Temperature units

°C

°C, °F, K

Fraction A/B max.

Application dependent

Application dependent

- Weight units

kg

mg, g, kg, ton, lb

- Time units

h

s, min, h, d

Totalizer 1

Mass flow

Mass flow, fraction A, fraction B, volume flow

Forward

Forward, reverse, net

Dim. dependent

mg, g, kg, ton, lb

Totalizer 2

Volume flow

Mass flow, fraction A, fraction B, volume flow

Forward

Forward, reverse, net

Dim. dependent

m

3

, ml, l, hl, kl, Ml, ft

3

, in

3

, US G, US MG,

UK G, UK MG

Low flow cut off

1.5 %

0 – 9.9 %

Empty pipe limit

Dim. dependent

-20000.0 g/cm

3

- +20000.0 g/cm

3

Noise filter

3

1 (min.) to 5 (max.)

Error level

Warning

Fatal, permanent, warning

Output

Current output 1

Off

Mass flow, fraction A, fraction B, volume flow,
sensor temp., density, pct. fraction A, off

- Direction

Unidirectional

Unidirectional, bidirectional

- Output mode

4 – 20 mA

0 - 20 mA, 4 - 20 mA

- Time constant

5 s

0 – 30 s

Digital output 1

Pulse

Pulse, frequency, error level, error number,
direction/limit, batch, off

Pulse

Mass flow

Mass flow, fraction A, fraction B, volume flow

- Amount/pulse

Dim. dependent

Dim. dependent

- Pulse polarity

Positive

Positive, negative

- Pulse width

66 ms

64 us, 130 us, 260us, 510 us, 1.0 ms, 2.0 ms,
4.1 ms, 8.2 ms, 16 ms, 33 ms, 66 ms, 130 ms,
260 ms, 520 ms, 1.0 s, 2.1 s, 4.2 s

Frequency

Off

Mass flow, fraction A, fraction B,
volume flow, sensor temp., density, pct.
fraction A

- Direction

Unidirectional

Unidirectional, bidirectional

- Max. frequency

10 kHz

500 Hz, 1 kHz, 5 kHz, 10 kHz

- Time constant

5 s

0 – 30 s

Error number

Off

0 - 255

7.7.1 Parameters