Commissioning
FlexCOR
™
Model CMF Series
70
Fluid Components Intl
Document 06EN003327 Rev. -
Parameter
Factory setting
Settings available
Password
Password
1000
1000 – 9999
Basic settings
Flow direction
Positive
Positive, negative
Massflow max.
Dim. dependent
Dim. dependent
- Weight units
Dim. dependent
mg, g, kg, ton, lb
- Time units
Dim. dependent
s, min, h, d
Volumeflow max.
Dim. dependent
Dim. dependent
- Volume units
Dim. dependent
m
3
, ml, l, hl, kl, Ml, ft
3
, in
3
, US G, US MG,
UK G, UK MG
- Time units
Dim. dependent
s, min, h, d
Density
- Minimum
+0.1 g/cm
3
-20000.0 kg/m
3
- +20000.0 kg/m
3
- Maximum
+2.0 g/cm
3
-20000.0 kg/m
3
- +20000.0 kg/m
3
- Weight units
g
mg, g, kg, ton, lb
- Time units
cm
3
cm
3
, m
3
, ft
3
, in
3
Sensor temp.
- Minimum
-50 °C
-250 °C - +250 °C
- Maximum
+250 °C
-250 °C - +250 °C
- Temperature units
°C
°C, °F, K
Fraction A/B max.
Application dependent
Application dependent
- Weight units
kg
mg, g, kg, ton, lb
- Time units
h
s, min, h, d
Totalizer 1
Mass flow
Mass flow, fraction A, fraction B, volume flow
Forward
Forward, reverse, net
Dim. dependent
mg, g, kg, ton, lb
Totalizer 2
Volume flow
Mass flow, fraction A, fraction B, volume flow
Forward
Forward, reverse, net
Dim. dependent
m
3
, ml, l, hl, kl, Ml, ft
3
, in
3
, US G, US MG,
UK G, UK MG
Low flow cut off
1.5 %
0 – 9.9 %
Empty pipe limit
Dim. dependent
-20000.0 g/cm
3
- +20000.0 g/cm
3
Noise filter
3
1 (min.) to 5 (max.)
Error level
Warning
Fatal, permanent, warning
Output
Current output 1
Off
Mass flow, fraction A, fraction B, volume flow,
sensor temp., density, pct. fraction A, off
- Direction
Unidirectional
Unidirectional, bidirectional
- Output mode
4 – 20 mA
0 - 20 mA, 4 - 20 mA
- Time constant
5 s
0 – 30 s
Digital output 1
Pulse
Pulse, frequency, error level, error number,
direction/limit, batch, off
Pulse
Mass flow
Mass flow, fraction A, fraction B, volume flow
- Amount/pulse
Dim. dependent
Dim. dependent
- Pulse polarity
Positive
Positive, negative
- Pulse width
66 ms
64 us, 130 us, 260us, 510 us, 1.0 ms, 2.0 ms,
4.1 ms, 8.2 ms, 16 ms, 33 ms, 66 ms, 130 ms,
260 ms, 520 ms, 1.0 s, 2.1 s, 4.2 s
Frequency
Off
Mass flow, fraction A, fraction B,
volume flow, sensor temp., density, pct.
fraction A
- Direction
Unidirectional
Unidirectional, bidirectional
- Max. frequency
10 kHz
500 Hz, 1 kHz, 5 kHz, 10 kHz
- Time constant
5 s
0 – 30 s
Error number
Off
0 - 255
7.7.1 Parameters