Commissioning

Fluid Components Intl

Document 06EN003327 Rev. -

61

FlexCor

Model CMF Series

Menu detail

7.5.4 Sensor

characteristics

When the SENSORPROM

®

memory unit is installed only the parameters correction factor,

density offset, fraction offset and table slope can be changed.

If you wish to change the mass flow measurement a percentage shift can be set under the menu
option correction factor. The change affects all flow-related values.

If you wish to change the density measurement a percentage shift can be set under the menu option

density factor.
If an offset needs to be added to the density measurement this is done with density offset.

If the fraction flowmeter function is used it is possible to customize the function in accordance with

the following equation:

Fraction = Ax + B, where

A = Table slope

B = Fraction offset

x = Measured fraction

Correction factors