Model cmf series 7.8.2 list of error numbers – Fluid Components International CMF Series Manual Electronics Setup and Menus User Manual
Page 23
![background image](/manuals/330625/23/background.png)
Commissioning
Fluid Components Intl
Document 06EN003327 Rev. -
73
FlexCOR
™
Model CMF Series
7.8.2 List of error
numbers
Error Error text
#Comment
Outputs
Input
No.
Remedy text
status
status
1
I1 - Power on
OK
Power on has activated
Active
Active
2
I2 - Add-on Module
Applied
A new module has been added to the system
Active
Active
3
I3 - Add-on Module
An add-on module is defective or has been removed.
Install
This can also be an internal add-on module
Active
Active
4
I4 - Param. corrected
A less vital parameter in the converter has been
OK
replaced by its default value
Active
Active
20
W20 - Totalizer 1
During initialization the check of the saved totalizer
Reset manually
value has failed. It is not possible to rely on the
saved totalizer value. The totalizer value must be
reset manually in order to rely on future readings
Active
Active
20
W20 - Totalizer 2
During initialization the check of the saved totalizer
Reset manually
value has failed. It is not possible to rely on the
saved totalizer. The totalizer value must be reset
manually in order to rely on future readings
Active
Active
21
W21 - Pulse overflow
Actual flow is too big compared with pulse width and
Reduced
Adjust pulse settings
mass/pulse
pulse width
Active
22
W22 - Batch timeout
Duration of batching has exceeded a predefined
Batch out-
Check installation
max. time
put on zero
Active
23
W23 - Batch overrun
Batch quantity has exceeded a predefined maximum
Batch out-
Check installation
overrun mass or volume
put on zero
Active
24
W24 - Batch neg. flow
Check flow direction
Negative flow direction during batch
Active
Active
30
W30 - Flowsaturated
Adjust max. flow
Flow is above Q
max.
settings
Max. 120 %
Active
31
W31- Empty pipe
Pipe is empty
Zero
Active
32
W32 - Temp. to high
The temperature of the fluid has exceeded the max.
Adjust temperature
temperature rating of the sensor (180 °C)
Active
Active
33
W33 - Temp. to low
The temperature of the fluid has exceeded the min.
Adjust temperature
temperature rating of the sensor (-50 °C)
Active
Active
34
W34 - Zero Adj. fail
The zero-point adjustment values are outside the limit
Check flow = zero
because there is not zero flow in the sensor. Check
zeroflow conditions, valves, pumps etc.
Active
Active
35
W35 - Current Out 1
Current output exceeds 120%. Ensure that the sensor is
Check max. settings
correctly sized and check max. flow setting
Active
Active
36
W36 - Freq/Pulse Out1
Freq/Pulse output exceeds 120%. Ensure that the sensor
Check max. settings
is correctly sized and check max. flow setting
Active
Active
40
P40 - SENSORPROM
®
Insert
SENSORPROM
®
unit not installed
Active
Active
41
P41 - Parameter range
A parameter is out of range.
Switch off and on
The error will disappear at the next power-on
Active
Active
42
P42 - Current output
Current loop is disconnected or the loop resistance
Check cables
is too big
Active
Active
43
P43 - Internal error
Internal error
Switch off and on
Active
Active
49
P49 - Protec. viol.
Too many errors occured at the same time.
Switch off and on
Some errors are not detected correctly
Active
Active
50
P50 - Temp. cable
Error in temperature sensor, check cables and
Check cable
connectors
Active
Active
51
P51 - Pick-up 1
Pick-up 1 amplitude too low. Check cables or application
Check cable/install.
for damping (air/gas in liquid)
Active
Active
52
P52 - Pick-up 2
Pick-up 2 amplitude too low. Check cables or application
Check cable/install.
for damping (air/gas in liquid)
Active
Active
60
F60 - CAN comm. error
CAN bus communication error. An add-on module, the
Converter/add-on module
display module or the converter is defective
Zero
Inactive
61
F61 - SENSORPROM
®
err.
It is not possible to rely on the data in SENSORPROM
®
Replace
unit
Active
Active
62
F62 - SENSORPROM
®
ID
The SENSORPROM
®
unit ID do not comply with the
Replace
product ID. The SENSORPROM
®
unit is from another
type of product.
Zero
Inactive
63
F63 - SENSORPROM
®
It is not possible to read from the SENSORPROM
®
Replace
unit
Active
Active
70
F70 - Pick-up phase
Check cables/polarity
Active
Active
71
F71 - Driver phase
Check cables/polarity
Active
Active
80-83
F80, 81, 82, 83 - Internal error Restart or replace
Active
Active
84
F84 - Sensor level
Pick-up amplitud saturated
Active
Active
97
F97 - Add-on module to old
Replace
Active
Active
Error code level:
W = Warning, F = Fatal, P = Permanent