Commissioning
Fluid Components Intl
Document 06EN003327 Rev. -
55
FlexCOR
™
Model CMF Series
Basic settings
Menu description
Setting of min./max.
values and units
Example; programming
of max. mass flow
Keypad operation
Implementation
Display on Transmitter
Push for 2 sec.
To access the
Password
user password
0000
Push once
To unlock password
CHANGE
0000
Push once
To enter 1000 as
CHANGE
password
1000
Push once
To lock password and to
CONV.SETUP MODE>
enter the menu
Basic settings
Push once
To enter basic setting
Flow direction
submenu
Positive
Push once
To go to mass flow
Massflow max.
max. setting
000020. kg/h
Push once
To change num.
Massflow max.
value
000020. kg/h
Push 4 times
To move the cursor
Massflow max.
to the num. position
000020. kg/h
Push
Until 4 appears
Massflow max.
000040. kg/h
Push once
To move the cursor to
Massflow max.
the next num. position
000040. kg/h
Push
Till 5 appears
Massflow max.
000045. kg/h
Push once
To move the cursor to
Massflow max.
the decimal point
000045. kg/h
Push
To position the decimal
Massflow max.
point correct
0000.45 kg/h
Push 3 times
To move cursor
Massflow max.
to "Kg" unit
0000.45 kg/h
Push twice
To change units to lb.
Massflow max.
0000.45Lb/h
Push once
To move cursor to
Massflow max.
the "h" unit
0000.45Lb/h
Push 3 times
To change "h" to "min"
Massflow max.
0000.45 Lb/min
Push
To lock the new setting
Massflow max.
of the mass flowmeter
000.45 Lb/m
Push twice
Transmitter reverts to
standard operation
The basic setting menu is used for basic configuration of the mass flow meter with a choice of units,
minimum and maximum limits for display and analog/digital outputs for all measurement
parameters, i.e. mass flow, volume flow, fraction, temperature and density.
Numerical values are entered by placing the cursor in the field that is to be set using the SELECT
key. Press unlock and the value can be changed using the change key. The desired value is locked
by activating LOCK.
Positioning of the decimal point is carried out by placing the cursor below the decimal point using
the SELECT key. The position can be set using the set key. The LOCK key is activated and the
decimal point is now positioned.
Selecting the unit: Place the cursor below the unit using SELECT key. Set the desired unit using
CHANGE key. Activate the LOCK key to save the setting. Place the cursor below the time scale
using SELECT key and choose the desired time scale using CHANGE, then save the desired value
by activating the LOCK key.
The maximum and minimum values set will then apply to all current and frequency/pulse outputs,
e.g. where the min. value will correspond to 0/4 mA depending on the setting of the current output
and the max. will correspond to 20 mA.
As example we want to change the default setting of the maximum mass flow on a CMF-A from
20 Kg/h to .45 lb/min.