Commissioning

Fluid Components Intl

Document 06EN003327 Rev. -

55

FlexCOR

Model CMF Series

Basic settings

Menu description

Setting of min./max.

values and units

Example; programming

of max. mass flow

Keypad operation

Implementation

Display on Transmitter

Push for 2 sec.

To access the

Password

user password

0000

Push once

To unlock password

CHANGE
0000

Push once

To enter 1000 as

CHANGE

password

1000

Push once

To lock password and to

CONV.SETUP MODE>

enter the menu

Basic settings

Push once

To enter basic setting

Flow direction

submenu

Positive

Push once

To go to mass flow

Massflow max.

max. setting

000020. kg/h

Push once

To change num.

Massflow max.

value

000020. kg/h

Push 4 times

To move the cursor

Massflow max.

to the num. position

000020. kg/h

Push

Until 4 appears

Massflow max.
000040. kg/h

Push once

To move the cursor to

Massflow max.

the next num. position

000040. kg/h

Push

Till 5 appears

Massflow max.
000045. kg/h

Push once

To move the cursor to

Massflow max.

the decimal point

000045. kg/h

Push

To position the decimal

Massflow max.

point correct

0000.45 kg/h

Push 3 times

To move cursor

Massflow max.

to "Kg" unit

0000.45 kg/h

Push twice

To change units to lb.

Massflow max.
0000.45Lb/h

Push once

To move cursor to

Massflow max.

the "h" unit

0000.45Lb/h

Push 3 times

To change "h" to "min"

Massflow max.
0000.45 Lb/min

Push

To lock the new setting

Massflow max.

of the mass flowmeter

000.45 Lb/m

Push twice

Transmitter reverts to
standard operation

The basic setting menu is used for basic configuration of the mass flow meter with a choice of units,

minimum and maximum limits for display and analog/digital outputs for all measurement

parameters, i.e. mass flow, volume flow, fraction, temperature and density.

Numerical values are entered by placing the cursor in the field that is to be set using the SELECT
key. Press unlock and the value can be changed using the change key. The desired value is locked

by activating LOCK.

Positioning of the decimal point is carried out by placing the cursor below the decimal point using

the SELECT key. The position can be set using the set key. The LOCK key is activated and the

decimal point is now positioned.

Selecting the unit: Place the cursor below the unit using SELECT key. Set the desired unit using
CHANGE key. Activate the LOCK key to save the setting. Place the cursor below the time scale

using SELECT key and choose the desired time scale using CHANGE, then save the desired value

by activating the LOCK key.

The maximum and minimum values set will then apply to all current and frequency/pulse outputs,

e.g. where the min. value will correspond to 0/4 mA depending on the setting of the current output
and the max. will correspond to 20 mA.

As example we want to change the default setting of the maximum mass flow on a CMF-A from
20 Kg/h to .45 lb/min.