beautypg.com

Brooks, Model 5850e – Brooks Instrument 5850E User Manual

Page 7

background image

1-1

Section 1 Introduction

Brooks

®

Model 5850E

Installation and Operation Manual
X-TMF-5850E-MFC-eng
Part Number: 541B102AAG
September, 2009

1-1 Purpose

The Brooks

®

Model 5850E Flow Controller is a mass flow measurement

device designed for accurately measuring and rapidly controlling flows of
gases. This instruction manual is intended to provide the user with all the
information necessary to install, operate and maintain the Brooks 5850E
Mass Flow Controller. This manual is organized into five sections:

Section 1

Introduction

Section 2

Installation

Section 3

Operation

Section 4

Maintenance & Troubleshooting

Section 5

Replacement Parts

Back Cover

Warranty, Local Sales/Service Contact Information

It is recommended that this manual be read in its entirety before attempting
to operate or repair the Model 5850E.

1-2 Description

The Brooks Model 5850E Mass Flow Controller is used widely in the
Semiconductor Industry as well as many others, where manual, electronic
or computer controlled gas handling occurs. The Model 5850E consists of
three basic units: a flow sensor, a control valve and an integral electronic
control system. This combination produces a stable gas flow, which
eliminates the need to continuously monitor and readjust gas pressures.
Standard features include:

Fast Response Control permits rapid gas settling times with little or no

over/undershoot. (Refer to Figure 1-1.)

Soft Start provides a flow ramping function which slows down the

introduction of the process gas for those processes which cannot
tolerate rapid flow transition. (Refer to Section 2-6 and Figure 1-2.)

Valve Override permits the user to fully open and close the control valve

independent of the command setting. (Refer to Section 2-8.)

Setpoint (Command) permits the user to program the mass flow controller

with an external 0-5 Vdc command potentiometer.
(Refer to Section 2-7.)

Low Command Valve Inhibit (Auto Shut-off) prevents the valve from

opening whenever the set-point is less than 2% of full scale.

Removable Cleanable Sensor permits the user to clean or replace the

sensor. (Refer to Section 4-4.)

Output Limiting prevents possible damage to delicate data acquisition

devices by limiting the output to +6.8 Vdc and -0.7 Vdc.