AC Voltage (Sine Wave) Specifications (cont.)
Absolute Uncertainty, tcal ± 5
°C
± (% of output +
µV)
Range Frequency
90 days
1 year
Resolution
Maximum
Burden
[1]
5725A Amplifier
45 Hz to 1 kHz
0.04
80 mV
0.05
80 mV
50 mA
1 to 20 kHz
0.06
100
0.08
100
70 mA
100 to 1020 V
20 to 30 kHz
0.08
100
0.10
100
70 mA
100 to 750 V
30 to 100 kHz
0.38
500
0.5
500
10 mV
70 mA
Auxiliary Output [dual output mode only]
[2]
10 to 20 Hz
0.15
370
0.2
370
20 to 45 Hz
0.08
370
0.1
370
45 Hz to 1 kHz
0.08
370
0.1
370
1 to 5 kHz
0.15
450
0.2
450
10 to 329.999 mV
5 to 10 kHz
0.3
450
0.4
450
1
µV 5
mA
10 to 20 Hz
0.15
450
0.2
450
20 to 45 Hz
0.08
450
0.1
450
45 Hz to 1 kHz
0.07
450
0.09
450
1 to 5 kHz
0.15
1400
0.2
1400
0.33 to 3.29999 V
5 to 10 kHz
0.3
1400
0.4
1400
10
µV 5
mA
[1]
Remote sensing is not provided. Output resistance is < 5 m
Ω for outputs ≥ 0.33 V. The AUX output resistance is < 1 Ω. The
maximum load capacitance is 500 pF, subject to the maximum burden current limits.