background image

12

Fluke Corporation

5500A Black and White Extended Specifications

Temperature Calibration (Thermocouple) Specifications

Absolute Uncertainty

Source/Measure,

tcal ± 5

°C, ± (°C)

[3]

Absolute Uncertainty

Source/Measure,

tcal ± 5

°C, ± (°C)

[3]

TC

Type

[1]

Range (

°C)

[2]

90 days

1 year

TC

Type

[1]

Range (

°C)

[2]

90 days

1 year

600 to 800

0.42

0.44

-200 to -100

0.37

0.37

800 to 1000

0.34

0.34

-100 to 800

0.26

0.26

1000 to 1550

0.30

0.30

L

800 to 900

0.17

0.17

B

1550 to 1820

0.26

0.33

-200 to -100

0.30

0.40

0 to 150

0.23

0.30

-100 to -25

0.17

0.22

150 to 650

0.19

0.26

-25 to 120

0.15

0.19

650 to 1000

0.23

0.31

120 to 410

0.14

0.18

1000 to 1800

0.38

0.50

N

410 to 1300

0.21

0.27

C

1800 to 2316

0.63

0.84

0 to 250

0.48

0.57

-250 to -100

0.38

0.50

250 to 400

0.28

0.35

-100 to -25

0.12

0.16

400 to 1000

0.26

0.33

-25 to 350

0.10

0.14

R

1000 to 1767

0.30

0.40

350 to 650

0.12

0.16

0 to 250

0.47

0.47

E

650 to 1000

0.16

0.21

250 to 1000

0.30

0.36

-210 to -100

0.20

0.27

1000 to 1400

0.28

0.37

-100 to -30

0.12

0.16

S

1400 to 1767

0.34

0.46

-30 to 150

0.10

0.14

-250 to -150

0.48

0.63

150 to 760

0.13

0.17

-150 to 0

0.18

0.24

J

760 to 1200

0.18

0.23

0 to 120

0.12

0.16

-200 to -100

0.25

0.33

T

120 to 400

0.10

0.14

-100 to -25

0.14

0.18

-200 to 0

0.56

0.56

-25 to 120

0.12

0.16

120 to 1000

0.19

0.26

K

1000 to 1372

0.30

0.40

U

0 to 600

0.27

0.27

The 10

µV/ °C linear output mode has the same uncertainty as the 300 mV dc range.

Applies to both simulated thermocouple output and thermocouple measurement.

[1]

Temperature standard ITS-90 or IPTS-68 is selectable.

[2] Resolution

is

0.01

°C.

[3]