Atec Fluke-5500A User Manual
Page 13
12
Fluke Corporation
5500A Black and White Extended Specifications
Temperature Calibration (Thermocouple) Specifications
Absolute Uncertainty
Source/Measure,
tcal ± 5
°C, ± (°C)
[3]
Absolute Uncertainty
Source/Measure,
tcal ± 5
°C, ± (°C)
[3]
TC
Type
[1]
Range (
°C)
[2]
90 days
1 year
TC
Type
[1]
Range (
°C)
[2]
90 days
1 year
600 to 800
0.42
0.44
-200 to -100
0.37
0.37
800 to 1000
0.34
0.34
-100 to 800
0.26
0.26
1000 to 1550
0.30
0.30
L
800 to 900
0.17
0.17
B
1550 to 1820
0.26
0.33
-200 to -100
0.30
0.40
0 to 150
0.23
0.30
-100 to -25
0.17
0.22
150 to 650
0.19
0.26
-25 to 120
0.15
0.19
650 to 1000
0.23
0.31
120 to 410
0.14
0.18
1000 to 1800
0.38
0.50
N
410 to 1300
0.21
0.27
C
1800 to 2316
0.63
0.84
0 to 250
0.48
0.57
-250 to -100
0.38
0.50
250 to 400
0.28
0.35
-100 to -25
0.12
0.16
400 to 1000
0.26
0.33
-25 to 350
0.10
0.14
R
1000 to 1767
0.30
0.40
350 to 650
0.12
0.16
0 to 250
0.47
0.47
E
650 to 1000
0.16
0.21
250 to 1000
0.30
0.36
-210 to -100
0.20
0.27
1000 to 1400
0.28
0.37
-100 to -30
0.12
0.16
S
1400 to 1767
0.34
0.46
-30 to 150
0.10
0.14
-250 to -150
0.48
0.63
150 to 760
0.13
0.17
-150 to 0
0.18
0.24
J
760 to 1200
0.18
0.23
0 to 120
0.12
0.16
-200 to -100
0.25
0.33
T
120 to 400
0.10
0.14
-100 to -25
0.14
0.18
-200 to 0
0.56
0.56
-25 to 120
0.12
0.16
120 to 1000
0.19
0.26
K
1000 to 1372
0.30
0.40
U
0 to 600
0.27
0.27
The 10
µV/ °C linear output mode has the same uncertainty as the 300 mV dc range.
Applies to both simulated thermocouple output and thermocouple measurement.
[1]
Temperature standard ITS-90 or IPTS-68 is selectable.
[2] Resolution
is
0.01
°C.
[3]
Does not include thermocouple error.