Additional specifications, Frequency specifications – Atec Fluke-5500A User Manual
Page 20
5500A Black and White Extended Specifications
Fluke Corporation
19
Additional Specifications
The following paragraphs provide additional specifications for the 5500A Calibrator ac voltage and ac current functions. These
specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5500A has been turned off. All extended range
specifications are based on performing the internal zero-cal function at weekly intervals, or when the ambient temperature changes by more
than 5
°C. (See Chapter 4, Front Panel Operations in the 5500A Operator Manual.)
Frequency Specifications
Frequency
Range
Resolution
1-Year Absolute Uncertainty,
tcal ± 5
°C
Jitter
0.01 - 119.99 Hz
0.01 Hz
25 ppm, ± 1 mHz
2
µs
120.0 - 1199.9 Hz
0.1 Hz
25 ppm, ± 1 mHz
2
µs
1.200 - 11.999 kHz
1.0 Hz
25 ppm, ± 1 mHz
[1]
2
µs
12.00 - 119.99 kHz
10 Hz
25 ppm, ± 15 mHz
140 ns
120.0 - 1199.9 kHz
100 Hz
25 ppm, ± 15 mHz
140 ns
1.200 - 2.000 MHz
1 kHz
25 ppm, ± 15 mHz
140 ns
[1]
± (25 ppm + 15 mHz) above 10 kHz