5500a specifications, Figure a. 5500a calibrator dimensional outline – Atec Fluke-5500A User Manual
Page 2
5500A Black and White Extended Specifications
Fluke Corporation
1
5500A Specifications
The following paragraphs detail specifications for the 5500A Calibrator. The specifications are valid after allowing a warm-up period of 30
minutes, or twice the time the 5500A has been turned off. For example, if the 5500A has been turned off for 5 minutes, the warm-up period
is 10 minutes.
All specifications apply for the temperature and time period indicated. For temperatures outside of tcal + 5
°C (tcal is the ambient
temperature when the 5500A was calibrated), the temperature coefficient is less than 0.1 times the 90-day specifications per
°C (limited to
0
°C to 50 °C). These specifications also assume the 5500A Calibrator is zeroed every seven days or when the ambient temperature
changes more than 5
°C. (See “Zeroing the Calibrator” in Chapter 4 of the 5500A Operator Manual.)
Also see additional specifications later in this chapter for information on extended specifications for ac voltage and current. The dimensional
outline for the 5500A Calibrator is shown in Figure A.
POWER
I
O
0
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
EARTH
SCOPE
BOOST
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
A
MULT
x
DIV
OUT
TRIG
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
43.2 cm (17 in)
47.0 cm (18.5 in)
6.4 cm
(2.5 in)
For Cable
Access
17.8 cm
(7 in)
om002f.ewps
Figure A. 5500A Calibrator Dimensional Outline