5500A Black and White Extended Specifications

Fluke Corporation

1

5500A Specifications

The following paragraphs detail specifications for the 5500A Calibrator. The specifications are valid after allowing a warm-up period of 30
minutes, or twice the time the 5500A has been turned off. For example, if the 5500A has been turned off for 5 minutes, the warm-up period
is 10 minutes.

All specifications apply for the temperature and time period indicated. For temperatures outside of tcal + 5

°C (tcal is the ambient

temperature when the 5500A was calibrated), the temperature coefficient is less than 0.1 times the 90-day specifications per

°C (limited to

0

°C to 50 °C). These specifications also assume the 5500A Calibrator is zeroed every seven days or when the ambient temperature

changes more than 5

°C. (See “Zeroing the Calibrator” in Chapter 4 of the 5500A Operator Manual.)

Also see additional specifications later in this chapter for information on extended specifications for ac voltage and current. The dimensional
outline for the 5500A Calibrator is shown in Figure A.

POWER

I

O

0

1

2

3

4

5

6

7

8

9

ENTER

M

k

m

V

Hz

FIELD

EDIT

/

+

F

OPR

EARTH

SCOPE

BOOST

MENU

PREV

SHIFT

RESET

CE

SETUP

REF

NEW

TC

MEAS

¡F

µ

n

p

W

dBm

sec

¡C

A

MULT

x

DIV

OUT

TRIG

5500A

CALIBRATOR

20V PK

MAX

HI

LO

TC

TRIG

OUT

1000V

RMS
MAX

20V
RMS
MAX

1V PK

MAX

20V PK

MAX

NORMAL

AUX

SCOPE

V, ,

RTD

A, -SENSE,

AUX V

200V PK

MAX

STBY

43.2 cm (17 in)

47.0 cm (18.5 in)

6.4 cm

(2.5 in)

For Cable

Access

17.8 cm

(7 in)

om002f.ewps

Figure A. 5500A Calibrator Dimensional Outline