5500A Black and White Extended Specifications

General Specifications

Warmup Time

Twice the time since last warmed up, to a maximum of 30 minutes.

Settling Time

Less than 5 seconds for all functions and ranges except as noted.

Standard Interfaces

IEEE-488 (GPIB), RS-232, 5725A Amplifier

Temperature Performance

• Operating: 0 °C to 50 °C
• Calibration (tcal): 15 °C to 35 °C
• Storage: -20 °C to 70 °C

Temperature Coefficient

Temperature Coefficient for temperatures outside tcal +5

°C is 0.1X/ °C of the 90-day

specification (or 1-year, as applicable) per °C.

Relative Humidity

[1]

• Operating: <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C
• Storage: <95 %, non-condensing

Altitude

• Operating: 3,050 m (10,000 ft) maximum
• Non-operating: 12,200 m (40,000 ft) maximum

Safety

Complies with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92

Analog Low Isolation

20 V

EMC

Designed to comply with FCC Rules Part 15; VFG 243/1991. If used in areas with
Electromagnetic fields of 1 to 3 V/m, resistance outputs have a floor adder of
0.508

Ω. Performance not specified above 3 V/m. This instrument may be susceptible

to electro-static discharge (ESD) from direct contact to the binding posts. Good static
aware practices should be followed when handling this and other pieces of
electronic equipment.

Line Power

• Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
• Line Frequency: 47 Hz to 63 Hz
• Line Voltage Variation: ±10 % about line voltage setting

Power Consumption

5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA

Dimensions

5500A Calibrator:
• Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on

bottom of unit

• Width, 43.2 cm (17 in), standard rack width
• Depth: 47.3 cm (18.6 in) overall 5725A Amplifier:
• Height, 13.3 cm (5.25 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on

bottom of unit

• Width, 43.2 cm (17 in), standard rack width
• Depth, 63.0 cm (24.8 in) overall.

Weight (without options)

5500A Calibrator, 22 kg (49 lb); 5725A Amplifier 32 kg (70 lb)

Absolute Uncertainty Definition

The 5500A specifications include stability, temperature coefficient, linearity, line and
load regulation, and the traceability of the external standards used for calibration.
You do not need to add anything to determine the total specification of the 5500A for
the temperature range indicated.

Specification Confidence
Interval

99 %

[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be required.