beautypg.com

Phase specifications – Atec Fluke-5500A User Manual

Page 18

background image

5500A Black and White Extended Specifications

Fluke Corporation

17

Phase Specifications

1-Year Absolute Uncertainty, tcal ± 5

°C, (

∆Φ

Degrees)

10 to 65 Hz

65 to 500 Hz

500 Hz to 1 kHz

1 to 5 kHz

5 to 10 kHz

0.15

°

[1]

0.9

°

[2]

2.0

°

[3]

6

° 10

°

[1]

For 33 to 1000 V output, burden current <6 mA. For 6 to 20 mA burden current (33 to 330 V), the phase uncertainty is 0.4 degree.

[2]

For 33 to 1000 V output, burden current <2 mA. For 2 to 5 mA burden current (33 to 330 V), the phase uncertainty is 1.5 degrees.

[3]

For 33 to 1000 V output, burden current <2 mA. For 2 to 5 mA burden current (33 to 330 V), the phase uncertainty is 5 degrees.

Power Uncertainty Adder due to Phase Error %

Phase (

Φ)

Watts

Degrees

Phase (

Φ)

VARs

Degrees

PF

10 to 65 Hz

65 to 500 Hz

500 Hz to 1

kHz

1 to 5 kHz

5 to 10 kHz

0 90

1.000

0.00 0.01

5 85

0.996

0.02 0.15

10 80

0.985

0.05 0.29

15 75

0.966

0.07 0.43

20 70

0.940

0.10 0.58

25 65

0.906

0.12 0.74

30 60

0.866

0.15 0.92

35 55

0.819

0.18 1.11

40 50

0.766

0.22 1.33

45 45

0.707

0.26 1.58

50 40

0.643

0.31 1.88

55 35

0.574

0.37 2.26

60 30

0.500

0.45 2.73

65 25

0.423

0.56 3.38

70 20

0.342

0.72 4.33

75 15

0.259

0.98 5.87

80 10

0.174

1.49 8.92

85 5

0.087

2.99

17.97

90 0

0.000

0.06 0.55

1.52

To calculate exact ac Watts power adders due to phase uncertainty for values not shown, use the following formula:

( )

( )

Adder

Cos

Cos

%

(

(

)

=

+

100 1

Φ ∆Φ

Φ

)

.

For example: for a PF of .9205 (

Φ = 23) and a phase uncertainty of

∆Φ

= 0.15, the ac Watts power adder is:

( )

( )

Adder

Cos

Cos

%

(

(23

)

=

+

=

100 1

15

23

0 11%

.

)

.

.