beautypg.com

Atec Fluke-5500A User Manual

Page 22

background image

5500A Black and White Extended Specifications

Fluke Corporation

21

AC Voltage (Sine Wave) Extended Bandwidth Specifications

1-Year Absolute Uncertainty,

tcal ± 5

°C,

± (% of output + % of range)

Range Frequency

% Output

% Range

Maximum Voltage Resolution

Normal Channel (Single Output Mode)

1.0 to 33 mV

Two digits, e.g., 25 mV

34 to 330 mV

Three digits

0.4 to 3.3 V

Two digits

4 to 33 V

0.01 to 10 Hz

5.0 %

0.5 %

Two digits

10 to 500 kHz

(See AC Voltage (Sine Waves) Specifications)

500 kHz to 1 MHz

-8 dB at 1 MHz, typical

0.3 to 3.3 V

1 to 2 MHz

-32 dB at 2 MHz, typical

Two digits

Auxiliary Output (Dual Output Mode)

10 to 330 mV

Three digits

0.4 to 3.3 V

0.01 to 10 Hz

5.0 %

0.5 %

Two digits

10 to 10 kHz

(See AC Voltage (Sine Wave) Specifications)