5500A Black and White Extended Specifications

Fluke Corporation

25

AC Current (Non-Sine Wave) Specifications

Triangle Wave &

Truncated Sine Wave

Ranges

[1]

Frequency

1-Year Absolute Uncertainty, tcal

± 5

°C,

± (% of output + % of range)

[2]

Maximum

Current Resolution

0.01 to 10 Hz

5.0 + 0.5

Two digits, e.g., 75 mA

10 to 45 Hz

0.25 + 0.5

45 Hz to 1 kHz

0.25 + 0.25

2.9 to 92.999 mA

1 to 10 kHz

0.25 + 0.5

Six digits on each range

0.01 to 10 Hz

5.0 + 0.5

10 to 45 Hz

0.25 + 0.5

Two digits

45 Hz to 1 kHz

0.25 + 0.5

93 to 929.999 mA

1 to 10 kHz

5.0 + 1.0

Six digits on each range

10 to 45 Hz

5.0 + 1.0

Two digits

45 Hz to 1 kHz

0.5 + 0.5

0.93 to 2.19 A

1 to 5 kHz

5.0 + 1.0

Six digits on each range

45 to 500 Hz

2.0 + 0.5

Two digits on each range

2.2 to 11 A

500 Hz to 1 kHz

5.0 + 1.0

Six digits on each range

0.01 to 10 Hz

5.0 + 0.5

10 to 45 Hz

0.25 + 0.5

Two digits, e.g., 50 mA

45 Hz to 1 kHz

0.25 +0.25

2.9 to 65.999 mA

1 to 10 kHz

0.25 + 0.5

Six digits on each range

0.01 to 10 Hz

5.0 + 0.5

Two digits

10 to 45 Hz

0.25 + 0.5

45 Hz to 1 kHz

0.25 + 0.5

66 to 659.999 mA

1 to 10 kHz

5.0 + 1.0

Six digits on each range

10 to 45 Hz

5.0 + 1.0

Two digits

45 Hz to 1 kHz

0.5 + 0.5

0.66 to 2.19 A

1 to 5 kHz

5.0 + 1.0

Six digits on each range

45 to 500 Hz

2.0 + 0.5

Two digits on each range

2.2 to 11 A

500 Hz to 1 kHz

5.0 + 1.0

Six digits on each range

[1]

All waveforms are p-p output ranges.

[2]

Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.

AC Current, Square Wave Characteristics (typical)

Range

Rise Time

Settling Time

Overshoot

I<4.4 A @ 400 Hz

25

µs 40

µs to 1 % of final value

<10 % for loads <100

AC Current, Triangle Wave Characteristics (typical)

Linearity to 400 Hz

Aberrations

0.3 % of p-p value, from 10 % to 90 % point

<1 % of p-p value, with amplitude >50 % of range