Electrical specifications, Dc voltage specifications – Atec Fluke-5500A User Manual
Page 4
5500A Black and White Extended Specifications
Fluke Corporation
3
Electrical Specifications
DC Voltage Specifications
Absolute Uncertainty, tcal ± 5
°C
± (% of output +
µV)
Stability
24 hours, ± 1
°C
Range
90 days
1 year
± (ppm output +
µV)
Resolution
µV
Maximum
Burden
[1]
0 to 329.9999 mV
0.005
3
0.006
3
5 ppm + 1
0.1
50
Ω
0 to 3.299999 V
0.004
5
0.005
5
4 + 3
1
10 mA
0 to 32.99999 V
0.004
50
0.005
50
4 + 30
10
10 mA
30 to 329.9999 V
0.004
500
0.0055
500
4.5 + 300
100
5 mA
100 to 1020.000 V
0.0045
1500
0.0055
1500
4.5 + 900
1000
5 mA
Auxiliary Output (dual output mode only)
[2]
0 to 329.999 mV
0.03
350
0.04
350
30 + 100
1
5 mA
0.33 to 3.3 V
0.03
350
0.04
350
30 + 100
10
5 mA
[1]
Remote sensing is not provided. Output resistance is < 5 m
Ω for outputs ≥ 0.33 V. The AUX output has an output resistance
of < 1
Ω.
[2]
Two channels of dc voltage output are provided.
Noise
Range
Bandwidth 0.1 to 10 Hz p-p
± (ppm output +
µV)
Bandwidth 10 to 10 kHz rms
0 to 329.9999 mV
1
µV 4
µV
0 to 3.299999 V
10
µV 50
µV
0 to 32.99999 V
100
µV 600
µV
30 to 329.9999 V
10 ppm + 1 mV
20 mV
100 to 1020.000 V
10 ppm + 5 mV
20 mV
Auxiliary Output (dual output mode only)
[1]
0 to 329.999 mV
5
µV 20
µV
0.33 to 3.3 V
20
µV 200
µV
[1]
Two channels of dc voltage output are provided.