5500A Black and White Extended Specifications

Fluke Corporation

3

Electrical Specifications

DC Voltage Specifications

Absolute Uncertainty, tcal ± 5

°C

± (% of output +

µV)

Stability

24 hours, ± 1

°C

Range

90 days

1 year

± (ppm output +

µV)

Resolution

µV

Maximum

Burden

[1]

0 to 329.9999 mV

0.005

3

0.006

3

5 ppm + 1

0.1

50

0 to 3.299999 V

0.004

5

0.005

5

4 + 3

1

10 mA

0 to 32.99999 V

0.004

50

0.005

50

4 + 30

10

10 mA

30 to 329.9999 V

0.004

500

0.0055

500

4.5 + 300

100

5 mA

100 to 1020.000 V

0.0045

1500

0.0055

1500

4.5 + 900

1000

5 mA

Auxiliary Output (dual output mode only)

[2]

0 to 329.999 mV

0.03

350

0.04

350

30 + 100

1

5 mA

0.33 to 3.3 V

0.03

350

0.04

350

30 + 100

10

5 mA

[1]

Remote sensing is not provided. Output resistance is < 5 m

Ω for outputs ≥ 0.33 V. The AUX output has an output resistance

of < 1

Ω.

[2]

Two channels of dc voltage output are provided.

Noise

Range

Bandwidth 0.1 to 10 Hz p-p

± (ppm output +

µV)

Bandwidth 10 to 10 kHz rms

0 to 329.9999 mV

1

µV 4

µV

0 to 3.299999 V

10

µV 50

µV

0 to 32.99999 V

100

µV 600

µV

30 to 329.9999 V

10 ppm + 1 mV

20 mV

100 to 1020.000 V

10 ppm + 5 mV

20 mV

Auxiliary Output (dual output mode only)

[1]

0 to 329.999 mV

5

µV 20

µV

0.33 to 3.3 V

20

µV 200

µV

[1]

Two channels of dc voltage output are provided.