Temperature calibration (rtd) specifications – Atec Fluke-5500A User Manual
Page 14
5500A Black and White Extended Specifications
Fluke Corporation
13
Temperature Calibration (RTD) Specifications
Absolute Uncertainty
tcal
±5 °C
± °C
[2]
Absolute Uncertainty
tcal
±5 °C
± °C
[2]
RTD Type
Range
°C
[1]
90 days
1 year
RTD Type
Range
°C
[1]
90 days
1 year
-200 to -80
0.04
0.05
-200 to -80
0.03
0.04
-80 to 0
0.05
0.05
-80 to 0
0.04
0.05
0 to 100
0.07
0.07
0 to 100
0.05
0.05
100 to 300
0.08
0.09
100 to 260
0.06
0.06
300 to 400
0.09
0.10
260 to 300
0.07
0.08
400 to 630
0.10
0.12
300 to 400
0.07
0.08
Pt 395,
100
Ω
630 to 800
0.21
0.23
400 to 600
0.08
0.09
-200 to -80
0.04
0.05
Pt 385,
500
Ω
600 to 630
0.09
0.11
-80 to 0
0.05
0.05
-200 to -80
0.03
0.03
0 to 100
0.07
0.07
-80 to 0
0.03
0.03
100 to 300
0.08
0.09
0 to 100
0.03
0.04
300 to 400
0.09
0.10
100 to 260
0.04
0.05
Pt 3926,
100
Ω
400 to 630
0.10
0.12
260 to 300
0.05
0.06
-200 to -190
0.25
0.25
300 to 400
0.05
0.07
-190 to -80
0.04
0.04
400 to 600
0.06
0.07
-80 to 0
0.05
0.05
Pt 385,
1000
Ω
600 to 630
0.22
0.23
0 to 100
0.06
0.06
-80 to 0
0.06
0.08
100 to 260
0.06
0.07
0 to 100
0.07
0.08
260 to 300
0.07
0.08
PtNi 385,
120
Ω
(Ni120)
100 to 260
0.13
0.14
300 to 400
0.08
0.09
400 to 600
0.08
0.10
Cu 427,
10
Ω
[3]
-100 to 260
0.3
0.3
Pt 3916,
100
Ω
600 to 630
0.21
0.23
-200 to -80
0.03
0.04
-80 to 0
0.03
0.04
0 to 100
0.04
0.04
100 to 260
0.04
0.05
260 to 300
0.11
0.12
300 to 400
0.12
0.13
400 to 600
0.12
0.14
Pt 385,
200
Ω
600 to 630
0.14
0.16
[1] Resolution
is
0.003
°C.
[2]
Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3]
Based on MINCO Application Aid No. 18.