Capacitance specifications – Atec Fluke-5500A User Manual
Page 12
5500A Black and White Extended Specifications
Fluke Corporation
11
Capacitance Specifications
Absolute Uncertainty, tcal ± 5
°C
± (% of output + nF)
Frequency
Range
90 days
1 year
Resolution
Allowed
Typical for <1 %
Error
0.33 to 0.4999 nF
0.38
0.01
0.5
0.01
0.1 pF
50 to 1000 Hz
10 kHz
0.5 to 1.0999 nF
0.38
0.01
0.5
0.01
0.1 pF
50 to 1000 Hz
10 kHz
1.1 to 3.2999 nF
0.38
0.01
0.5
0.01
0.1 pF
50 to 1000 Hz
10 kHz
3.3 to 10.999 nF
0.38
0.01
0.5
0.01
1 pF
50 to 1000 Hz
10 kHz
11 to 32.999 nF
0.19
0.1
0.25
0.1
1 pF
50 to 1000 Hz
10 kHz
33 to 109.99 nF
0.19
0.1
0.25
0.1
10 pF
50 to 1000 Hz
10 kHz
110 to 329.99 nF
0.19
0.3
0.25
0.3
10 pF
50 to 1000 Hz
10 kHz
0.33 to 1.0999
µF
0.19
1
0.25
1
100 pF
50 to 1000 Hz
5 kHz
1.1 to 3.2999
µF
0.26
3
0.35
3
100 pF
50 to 1000 Hz
2 kHz
3.3 to 10.999
µF
0.26
10
0.35
10
1 nF
50 to 400 Hz
1.5 kHz
11 to 32.999
µF
0.30
30
0.40
30
1 nF
50 to 400 Hz
800 Hz
33 to 109.99
µF
0.38
100
0.50
100
10 nF
50 to 200 Hz
400 Hz
110 to 329.99
µF
0.50
300
0.70
300
10 nF
50 to 100 Hz
200 Hz
330 to 1.1 mF
1
300
1
300
100 nF
50 to 100 Hz
150 Hz
Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters.
The output is continuously variable from 330 pF to 1.1 mF.
For all ranges, the maximum charge and discharge current is 150 mA pk or 30 mA rms. The peak voltage is 4 V, except the 330
µF to
1.1 mF range is limited to 1 V. The maximum lead resistance for no additional error in 2-wire COMP mode is 10
Ω.