beautypg.com

Ac voltage, square wave characteristics – Atec Fluke-5500A User Manual

Page 25

background image

24

Fluke Corporation

5500A Black and White Extended Specifications

AC Voltage, Square Wave Characteristics

Rise Time

@ 1 kHz

Typical

Settling Time

@ 1 kHz

Typical

Overshoot

@ 1 kHz

Typical

Duty Cycle Range

Duty Cycle Uncertainty

[1]

<1

µs

<10

µs to 1 %

of final value

<2 %

1 % to 99 %, <3.3 V p-p,

0.01 Hz to 100 kHz

± (0.8 % of period +140 ns) for

frequencies >10 kHz; + (0.8 % of

period + 2

µs) for frequencies

≤10 kHz.

[1]

For duty cycles of 10.00 % to 90.00 %.

AC Voltage, Triangle Wave Characteristics (typical)

Linearity to 1 kHz

Aberrations

0.3 % of p-p value, from 10 % to 90 % point

<1 % of p-p value, with amplitude >50 % of range

AC Current (Sine Wave) Extended Bandwidth Specifications

1-Year Absolute Uncertainty,

tcal ± 5

°C,

± (% of output + % of range)

Range Frequency

% Output

% Range

Maximum

Current Resolution

All current ranges, <330 mA

0.01 to 10 Hz

5.0

0.5

2 digits each range

10 to 10 kHz

(See AC Current (Sine Wave) Specifications)