beautypg.com

Dc current specifications – Atec Fluke-5500A User Manual

Page 5

background image

4

Fluke Corporation

5500A Black and White Extended Specifications

DC Current Specifications

Absolute Uncertainty,

tcal ± 5

°C

± (% of output +

µA)

Range

90 days

1 year

Resolution

Compliance

Voltage

Maximum

Inductive

Load

0 to 3.29999 mA

0.010

0.05

0.013

0.05

0.01

µA 4.5

V 1

µH

0 to 32.9999 mA

0.008

0.25

0.01

0.25

0.1

µA 4.5

V 200

µH

0 to 329.999 mA

0.008

3.3

0.01

3.3

1

µA

4.5 to 3.0 V

[1]

200

µH

0 to 2.19999 A

0.023

44

0.03

44

10

µA

4.5 to 3.4 V

[2]

200

µH

0 to 11 A

0.038

330

0.06

330

100

µA

4.5 to 2.5 V

[3]

200

µH

5725A Amplifier

0 to 11 A

0.03

330

0.04

330

100

4 V

400

µH

[1]

The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -5.05*Io+4.67. The highest compliance voltage is limited to 4.5 V.

[2]

The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.588*Io+4.69. The highest compliance voltage is limited to 4.5 V.

[3]

The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.204*Io+4.75. The highest compliance voltage is limited to 4.3 V.

Noise

Ranges

Bandwidth

0.1 to 10 Hz

p-p

Bandwidth

10 to 10 kHz

rms

0 to 3.29999 mA

20 nA

200 nA

0 to 32.9999 mA

200 nA

2.0

µA

0 to 329.999 mA

2000 nA

20

µA

0 to 2.19999 A

20

µA 1

mA

0 to 11 A

200

µA 10

mA

5725A Amplifier

0 to 11 A

± 25 ppm of output + 200 nA

2 mA