Dc current specifications – Atec Fluke-5500A User Manual
Page 5
4
Fluke Corporation
5500A Black and White Extended Specifications
DC Current Specifications
Absolute Uncertainty,
tcal ± 5
°C
± (% of output +
µA)
Range
90 days
1 year
Resolution
Compliance
Voltage
Maximum
Inductive
Load
0 to 3.29999 mA
0.010
0.05
0.013
0.05
0.01
µA 4.5
V 1
µH
0 to 32.9999 mA
0.008
0.25
0.01
0.25
0.1
µA 4.5
V 200
µH
0 to 329.999 mA
0.008
3.3
0.01
3.3
1
µA
4.5 to 3.0 V
[1]
200
µH
0 to 2.19999 A
0.023
44
0.03
44
10
µA
4.5 to 3.4 V
[2]
200
µH
0 to 11 A
0.038
330
0.06
330
100
µA
4.5 to 2.5 V
[3]
200
µH
5725A Amplifier
0 to 11 A
0.03
330
0.04
330
100
4 V
400
µH
[1]
The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -5.05*Io+4.67. The highest compliance voltage is limited to 4.5 V.
[2]
The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.588*Io+4.69. The highest compliance voltage is limited to 4.5 V.
[3]
The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.204*Io+4.75. The highest compliance voltage is limited to 4.3 V.
Noise
Ranges
Bandwidth
0.1 to 10 Hz
p-p
Bandwidth
10 to 10 kHz
rms
0 to 3.29999 mA
20 nA
200 nA
0 to 32.9999 mA
200 nA
2.0
µA
0 to 329.999 mA
2000 nA
20
µA
0 to 2.19999 A
20
µA 1
mA
0 to 11 A
200
µA 10
mA
5725A Amplifier
0 to 11 A
± 25 ppm of output + 200 nA
2 mA