5500A Black and White Extended Specifications

Fluke Corporation

5

Resistance Specifications

Absolute Uncertainty, tcal ± 5

°C

± (% of output +

)

[2]

Range

[1]

90 days

1 year

Resolution

Allowable

Current

[4]

0 to 10.99

Ω 0.009

0.008

[3]

0.012 0.008

[3]

0.001

1 to 125 mA

11 to 32.999

Ω 0.009

0.015

[3]

0.012 0.015

[3]

0.001

1 to 125 mA

33 to 109.999

Ω 0.007 0.015

[3]

0.009 0.015

[3]

0.001

1 to 70 mA

110 to 329.999

Ω 0.007 0.015

[3]

0.009 0.015

[3]

0.001

1 to 40 mA

330

Ω to 1.09999 kΩ 0.007

0.06

0.009

0.06

0.01

250

µA to 18 mA

1.1 to 3.29999 k

Ω 0.007 0.06 0.009 0.06

0.01

250

µA to 5 mA

3.3 to 10.9999 k

Ω 0.007 0.6 0.009 0.6

0.1

25

µA to 1.8 mA

11 to 32.9999 k

Ω 0.007 0.6 0.009 0.6

0.1

25

µA to 0.5 mA

33 to 109.999 k

0.008 6 0.011 6

1 2.5

µA to 0.18 mA

110 to 329.999 k

0.009 6 0.012 6

1 2.5

µA to 0.05 mA

330 k

Ω to 1.09999 MΩ

0.011

55

0.015

55

10

250 nA to 0.018 mA

1.1 to 3.29999 M

0.011

55

0.015

55

10

250 nA to 5

µA

3.3 to 10.9999 M

0.045

550

0.06

550

100

25 nA to 1.8

µA

11 to 32.9999 M

0.075

550

0.1

550

100

25 nA to 0.5

µA

33 to 109.999 M

0.4

5500

0.5

5500

1000

2.5 nA to 0.18

µA

110 to 330 M

0.4

16500

0.5

16500

1000

2.5 nA to 0.06

µA

[1]

Continuously variable from 0 to 330 M

Ω.

[2]

Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.

[3]

The floor adder is improved to 0.006

Ω (0 to 10.99 Ω range) and 0.010 Ω (11 to 329.999 Ω) if the 5500A Calibrator is zeroed

(ohms zero or instrument zero) within 8 hours and temperature is ±1

°C of zeroing ambient temperature.

[4]

Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases by Floor(new) =
Floor(old) x Imin/Iactual. For example, a 100

µA stimulus measuring 100 Ω has a floor uncertainty of 0.01 Ω x 1 mA/100 µA

= 0.1

Ω.

Range Maximum

Voltage

[1]

Maximum

Lead

Resistance

[2]

0 to 10.99

Ω 1.37

<3.2

11 to 32.999

Ω 4.12

<3.2

33 to 109.999

Ω 7.7

<3.2

110 to 329.999

Ω 13.2

<3.2

330

Ω to 1.09999 kΩ 19.8

<6

1.1 to 3.29999 k

Ω 16.5

<6

3.3 to 10.9999 k

Ω 19.8

<6

11 to 32.9999 k

Ω 16.5

<6

33 to 109.999 k

Ω 19.8

<6

110 to 329.999 k

Ω 16.5

330 k

Ω to 1.09999 MΩ 19.8

1.1 to 3.29999 M

Ω 16.5

3.3 to 10.9999 M

Ω 19.8

11 to 32.9999 M

Ω 16.5

33 to 109.999 M

Ω 19.8

110 to 330 M

Ω 19.8

(n/a 110 k

Ω and above)

[1]

This is for the largest resistance for each range. The maximum voltage for other values is Imax (highest value of Allowable Current
above) multiplied by Rout.

[2]

Maximum lead resistance for no additional error in 2-wire COMP.