5500A Black and White Extended Specifications

Fluke Corporation

15

AC Power (45 Hz to 65 Hz) Specification Summary, PF=1

Absolute Uncertainty, tcal ± 5

°C, ± (% of Watts output)

[1]

Current Range

Voltage Range

3.3 to 8.999 mA

9 to 32.999 mA

33 to 89.99 mA

90 to 329.99 mA

5500A Calibrator

33 to 329.999 mV

0.30

0.20

0.25

0.20

90 days

330 mV to 1020 V

0.20

0.12

0.20

0.12

33 to 329.999 mV

0.40

0.25

0.35

0.25

1 year

330 mV to 1020 V

0.25

0.15

0.25

0.15

5725A Amplifier

90 days

100 to 1020 V

0.20

0.12

0.20

0.12

1 year

100 to 1020 V

0.25

0.15

0.25

0.15

0.33 to 0.8999 A

0.9 to 2.1999 A

2.2 to 4.4999 A

4.5 to 11 A

5500A Calibrator

33 to 329.999 mV

0.25

0.20

0.25

0.20

90 days

330 mV to 1020 V

0.20

0.12

0.18

0.12

33 to 329.999 mV

0.35

0.25

0.35

0.25

1 year

330 mV to 1020 V

0.25

0.15

0.20

0.15

5725A Amplifier

90 days

100 to 1020 V

0.20

0.12

0.18

0.12

1 year

100 to 1020 V

0.25

0.15

0.20

0.15

1.5 to 4.4999 A

4.5 to 11 A

5500A Calibrator

33 to 329.999 mV

0.25

0.20

90 days

330 mV to 1020 V

0.15

0.12

33 mV to 1020 V

0.35

0.25

1 year

330 mV to 1020 V

0.20

0.15

[1]

To determine uncertainty with more precision, see “Calculating Power Uncertainty.”