Level

standard

–145 dBm to +20 dBm

Setting range

with R&S

®

SMU-B31 or R&S

®

SMU-B36

option

to +30 dBm

standard

f

≤ 3 GHz

+13 dBm (PEP)

2

f > 3 GHz

+11 dBm (PEP)

with R&S

®

SMU-B31 or R&S

®

SMU-B36

option

f

≤ 3 GHz

+19 dBm (PEP)

Maximum level

f > 3 GHz

+17 dBm (PEP)

Resolution

0.01

dB

for levels > –120 dBm,
attenuator mode: auto,
temperature range = +18 °C to +28 °C

1 MHz

≤ f ≤ 3 GHz

< 0.5 dB

Level uncertainty

f > 3 GHz

< 0.9 dB

Additional uncertainty with ALC state: off
(sample & hold)

This function is needed only for some
special applications.

< 0.2 dB

standard, ALC state: on

f

≤ 3 GHz

< 1.6, < 1.4 (typ.)

f > 3 GHz

< 1.85, < 1.6 (typ.)

with R&S

®

SMU-B31 or R&S

®

SMU-B36

option, ALC state: on


attenuator mode: normal

f

≤ 3 GHz

< 1.65, < 1.45 (typ.)

f > 3 GHz

< 1.9, < 1.65 (typ.)

attenuator mode: high power

f

≤ 3 GHz

< 1.7, < 1.5 (typ.)

Output impedance
VSWR in 50

Ω system

f > 3 GHz

< 1.9, < 1.65 (typ.)

after IEC/IEEE bus delimiter,
to < 0.1 dB deviation from final value,
with GUI update stopped,
temperature range = +18 °C to +28 °C

ALC state: on

< 2 ms, 1.5 ms (typ.)

ALC state: off

< 4 ms, 2.5 ms (typ.)

in list mode after trigger impulse to 0.3 dB
deviation from final value

< 450 µs, 300 µs (typ.)

Setting time

1

range switchover with R&S

®

SMU-B31 or

R&S

®

SMU-B36 option

< 10 ms

attenuator mode: fixed, ALC state: on

Uninterrupted level setting

setting range

> 20 dB

maximum permissible RF power in output
frequency range for f > 1 MHz

1 MHz

≤ f ≤ 3 GHz

50 W

f > 3 GHz

10 W

Back-feed (from

≥ 50 Ω source)

maximum permissible DC voltage

50 V

2

PEP = peak envelope power.