Analog modulation, Internal modulation generator, Input for external modulation signals – Atec Rohde-Schwarz-SMU200A User Manual
Page 16: Amplitude modulation, Wideband amplitude modulation
Version 08.00, January 2012
16
Rohde & Schwarz R&S
®
SMU200A Vector Signal Generator
Analog modulation
Internal modulation generator
Frequency range
0.1 Hz to 1 MHz
Resolution of setting
0.1 Hz
Frequency uncertainty
< 0.012 Hz
+
relative deviation of reference frequency
up to 100 kHz
< 0.1 dB
Frequency response
up to 1 MHz
< 1 dB
Distortion
up to 100 kHz at R
L
> 200
Ω,
level = 1 V (V
p
)
< 0.1 %
V
p
at LF connector, R
L
> 200
Ω
1 mV to 3 V
resolution 1
mV
Output voltage
setting uncertainty at 1 kHz
< (1 % of reading + 1 mV)
Output impedance
16
Ω
Frequency setting time
to within < 1 × 10
−7
, with GUI update
stopped, after IEC/IEEE bus delimiter
< 3 ms
digital sweep in discrete steps
operating modes
automatic, step, single,
external single, external step,
manual or external trigger,
linear or logarithmic spacing
sweep range
entire frequency range
linear step width
entire frequency range
Sweep
logarithmic step width
0.01 % to 100 % per step
Input for external modulation signals
input impedance
high (> 100 k
Ω),
switchable to 50
Ω with R&S
®
SMU-B20 or
R&S
®
SMU-B22 option
input sensitivity (peak value for set
modulation depth or deviation)
1 V
Modulation input
EXT MOD
maximum permissible input voltage
±10 V
Amplitude modulation
Operating modes
internal, external AC/DC
Modulation depth
modulation is clipped at high levels if
maximum PEP is reached
0 % to 100 %
Resolution
0.1
%
Setting uncertainty
attenuator mode: auto,
f
mod
= 1 kHz and m < 80 %
< (1 % of reading + 1 %)
PEP in specified range,
attenuator mode: auto
f
≤ 3 GHz, at f
mod
= 1 kHz,
m = 30 %
< 0.5 %
m = 80 %
< 0.8 %
f > 3 GHz, at f
mod
= 1 kHz,
m = 30 %
< 1 %
AM distortion
m = 80 %
< 1.6 %
Modulation frequency range
DC, 20 Hz to 500 kHz
Modulation frequency response
AC mode, 20 Hz to 500 kHz
< 1 dB
Synchronous
φM at AM
m = 30 %, f
mod
= 1 kHz, peak value
< 0.1 rad
Wideband amplitude modulation
Operating modes
modulation input I
external DC
Modulation frequency response
as with I/Q modulation – external wideband I/Q
Input impedance
50
Ω
Input sensitivity
peak voltage for 100 % AM
0.25 V