beautypg.com

Atec Rohde-Schwarz-SMU200A User Manual

Page 23

background image

Version 08.00, January 2012

Rohde & Schwarz

R&S

®

SMU200A Vector Signal Generator

23

Carrier leakage

referenced to full scale

<

−55 dBc, < −65 dBc (typ.)

up to 10 MHz

56 dB (typ.)

Suppression of image sideband for entire
instrument in modulation bandwidth

up to 30 MHz

50 dB (typ.)

wideband noise, with full-scale DC input

20 MHz

≤ f ≤ 200 MHz

–145 dBc (typ.)

200 MHz < f

≤ 1.5 GHz

–148 dBc (typ.)

1.5 GHz < f

≤ 3 GHz

–148 dBc (typ.)

f > 1.5 GHz

–145 dBc (typ.)

ACLR with an ideal input signal, 3GPP test model 1, 64 DPCH
level

≤ 10.5 dBm PEP, ≤ 16.5 dBm PEP with R&S

®

SMU-B31, R&S

®

SMU-B36 options

frequency = 1800 MHz to 2200 MHz

5 MHz offset (baseband gain: 3 dB)

70 dB (typ.)

RF spectral purity

10 MHz offset (baseband gain: 6 dB)

73 dB (typ.)

Digital I/Q inputs

standard

in line with Rohde & Schwarz TVR290,
I/Q data and control signals, data and
interface clock

level LVDS
connector 26-pin

MDR

Interface

data rate

66 MHz to 100 MHz

source

user-defined, digital I/Q in

sample rate

400 Hz to 100 MHz

resolution (user-defined)

0.001 Hz

I/Q sample rate
(max. sample rate depends on interface
data rate)

With source ‘user-defined’, the sample
rate must be entered via the parameter
‘sample rate’, no I/Q data clock being
necessary. With source ‘digital I/Q in’, the
sample rate will be estimated on the basis
of the applied I/Q data clock.

frequency uncertainty (user-defined)

< 5 × 10

14

resolution 18

bit

logic format

two’s complement

bandwidth

sample rate = 100 MHz
(no interpolation, user-defined)

40 MHz

I/Q data

sample rate < 100 MHz (interpolation)

0.31 × sample rate

markers 4

Control signals

data valid

valid samples marked in data stream