beautypg.com

Rf characteristics, Frequency, Frequency sweep – Atec RohdeSchwarz-SMW200A User Manual

Page 7: Reference frequency

background image

Version 01.02, September 2013

Rohde & Schwarz R&S

®

SMW200A Vector Signal Generator

7

RF characteristics

Frequency

Range R&S

®

SMW-B103, R&S

®

SMW-B203

100 kHz to 3 GHz

R&S

®

SMW-B106, R&S

®

SMW-B206

100 kHz to 6 GHz

Resolution of setting

0.001 Hz

Resolution of synthesis

fundamental frequency range = 750 MHz to 1500 MHz
standard

5 µHz (nom.)

with R&S

®

SMW-B22 option

0.2 µHz (nom.)

Setting time

1

to within < 1 × 10

–7

for f > 200 MHz or < 124 Hz for f < 200 MHz,

with GUI update stopped
after IEC/IEEE bus delimiter

< 1.2 ms, 0.6 ms (typ.)

Resolution of phase offset setting

0.1°

Frequency sweep

Operating mode

digital sweep in discrete steps

Trigger modes

free run

auto

execute one full sweep

single

execute one step

step

sweep start and stop controlled by
external trigger signal

start/stop

Trigger source

external trigger signal (INST TRG A or B
at rear), rotary knob, touchpanel, remote
control

Sweep range

full frequency range

Sweep shape

sawtooth, triangle

Step size

linear

full frequency range

logarithmic

0.01 % to 100 % per step

Dwell time setting range

1 ms to 100 s

Dwell time setting resolution

0.1 ms

Reference frequency

Frequency error

at time of calibration in production

standard

< 1 × 10

–8

with R&S

®

SMW-B22 option

< 5 × 10

–9

Aging

after 30 days of uninterrupted operation

standard

1 × 10

–9

/day, 1 × 10

–7

/year

with R&S

®

SMW-B22 option

5 × 10

–10

/day, 3 × 10

–8

/year

Temperature effect

in temperature range from 0 °C to +50 °C

standard

6 × 10

–8

with R&S

®

SMW-B22 option

6 × 10

–9

Warm-up time

to nominal thermostat temperature

≤ 10 min

Output for internal reference frequency
Connector type

REF OUT on rear panel

BNC female

Output frequency

sine wave

10 MHz or external input frequency

Output level

2 dBm to 8 dBm,
5 dBm to 7 dBm (typ.)

Source impedance

50 Ω (nom.)

Input for external reference frequency
Connector type

REF IN on rear panel

BNC female

Input frequency

5 MHz, 10 MHz or 13 MHz

Min. frequency locking range

standard

±3 × 10

–6

with R&S

®

SMW-B22 option

±1.5 × 10

–7

Input level range

level limits

≥ –6 dBm, ≤ 19 dBm

recommended input level

0 dBm to 19 dBm

Input impedance

50 Ω (nom.)

1

Installation of software that is not authorized by Rohde & Schwarz for use on the R&S

®

SMW200A or installation of antivirus software can deteriorate