beautypg.com

3gpp fdd (with r&s, Smw-k42 option) – Atec RohdeSchwarz-SMW200A User Manual

Page 29

background image

Version 01.02, September 2013

Rohde & Schwarz R&S

®

SMW200A Vector Signal Generator

29

Modulation performance for digital standards and modulation systems

3GPP FDD (with R&S

®

SMW-K42 option)

Error vector magnitude

1 DPCH, RMS,
frequency = 1800 MHz to 2200 MHz

< 0.8 %, 0.3 % (meas.)

Adjacent channel leakage ratio (ACLR)

test model 1, 64 DPCHs, frequency = 1800 MHz to 2200 MHz,
average channel power ≤ 5 dBm

5 MHz offset

> 70 dB

10 MHz offset

> 72 dB

Measured ACPR for 3GPP test model 1, 64 DPCH.

Measured ACPR for a 3GPP four-carrier signal with test model 1, 64 DPCH on each carrier.