Dynamic fading (r&s, Smw-k71 option) – Atec RohdeSchwarz-SMW200A User Manual
Page 34
Version 01.02, September 2013
34
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Predefined settings
standard
LTE, GSM, CDMA2000
®
, 1xEV-DO,
WLAN, WiMAX™ ITU, NADC, PCN,
TETRA
with R&S
®
SMW-K71 option
3GPP FDD WCDMA
with R&S
®
SMW-K72 option
WiMAX™ SUI, DAB
with R&S
®
SMW-K74 option
LTE MIMO, IEEE 802.11n MIMO,
IEEE 802.11ac MIMO, WiMAX™ MIMO
Dynamic fading (R&S
®
SMW-K71 option)
At least one R&S
®
SMW-B14 fading simulator must be installed. If two or more R&S
®
SMW-B14 are installed (signal paths A and B),
dynamic fading functions can be used either on signal path A or B with one R&S
®
SMW-K71 option. For dynamic fading functions to be
used on signal paths A and B simultaneously, two R&S
®
SMW-K71 must be installed.
Moving delay mode
Number of fading paths
2 per signal path
Fading profiles
none
Basic delay
in steps of 5 ns
0 s to 0.5 s
Delay variation
peak to peak
0.3 µs to 40 µs
variation period
10 s to 500 s
variation speed
0 µs/s to 5 µs/s
Delay step size
5 ps
Birth-death mode
System bandwidth
160 MHz
Number of fading paths
2 per signal path
Fading profiles
pure Doppler
Delay range
0 s to 40 µs
Delay grid
0 s to 20 µs
11
Positions
3 to 50
11
Hopping dwell
100 ms to 5 s
Start offset
separately settable for each signal path
1 ms to 200 ms
Delay resolution
10 ns
High-speed train
Fading profiles
static path, pure Doppler, Rayleigh
Speed
at f = 1 GHz
0 km/h to 4320 km/h
D (min)
1 m to 100 m
D (s)
20 m to 2000 m
Two-channel interferer
Number of fading paths
2 per signal path
Fading profiles
static path, pure Doppler, Rayleigh,
Fading profile parameter
Rayleigh
pseudo-noise interval
> 1 year
phase resolution
1°
Pure Doppler
maximum resulting Doppler shift
frequency ratio × current Doppler
frequency
frequency ratio
–1 to +1
resolution 0.01
Fading path loss
setting range
0 dB to 50 dB
resolution 0.01
dB
accuracy
< 0.01 dB
Speed range
at f = 1 GHz
0 km/h to 4320 km/h
accuracy
< 0.1 %
Min. delay
path 1
0 µs to 1638 µs
path 2
0 µs to 999.9 µs
Max. delay
path 1
n/a
path 2
0.1 µs to 1000 µs
Moving mode
path 1
n.a.
path 2
sliding, hopping
Period/dwell
0.1 s to 10 s
11