Envelope tracking (r&s, Smw-k540 option) – Atec RohdeSchwarz-SMW200A User Manual
Page 22
Version 01.02, September 2013
22
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Bias voltage (single-ended and differential) EMF
–3.6 V to +3.6 V
7
Resolution
2
mV
Uncertainty
1 % + 4 mV
Offset voltage
Differential
EMF
–300 mV to +300 mV
Resolution
0.1
mV
Uncertainty
1 % + 0.1 % × bias voltage + 1 mV
Differential signal balance
at R
L
= 50 Ω, output voltage > 0.5 V (V
p
)
Magnitude
up to 10 MHz
< 0.2 dB, 0.05 dB (meas.)
up to 80 MHz
0.2 dB (meas.)
Frequency response
8
at
R
L
= 50 Ω, output voltage > 0.5 V (V
p
)
Magnitude
up to 10 MHz
0.02 dB (meas.)
up to 80 MHz
0.03 dB (meas.)
Envelope tracking (R&S
®
SMW-K540 option)
With this option, the analog I/Q outputs can be used to generate an analog signal corresponding to the envelope of the I/Q signal to
test envelope tracking modulators..
This option can be installed once if the instrument is equipped with the R&S
®
SMW-B13 option. If the instrument is equipped with the
R&S
®
SMW-B13T option, envelope tracking can be used either on signal path A or B with one R&S
®
SMW-K540 option. For envelope
tracking to be used on signal paths A and B simultaneously, two R&S
®
SMW-K540 must be installed.
For each R&S
®
SMW-K540 option to be installed, a R&S
®
SMW-K16 option must be installed, and the instrument must be equipped
with at least one baseband generator (R&S
®
SMW-B10 option)
General
Envelope voltage adaption
auto, manual
Output type
single-ended, differential
Bias voltage
see section “Differential analog I/Q outputs”
Offset voltage
see section “Differential analog I/Q outputs”
Envelope to RF delay
Setting range
–500 ns to +500 ns
Setting resolution
1 ps
Shaping
off, linear, from table, polynomial,
dethroughing
Envelope voltage adaption mode auto
Power Amplifier Input Power P
in
Setting range
–145.00 dB to +30.00 dB
Setting resolution
0.01 dB
Power Amplifier Supply Voltage V
cc
see “Output voltage “ in section “Differential analog I/Q outputs”
DC Modulator gain
0.00 dB to +10.00 dB
Envelope voltage adaption mode manual
Pre-gain
Setting range
–20.00 dB to 0.00 dB
Setting resolution
0.01 dB
Post-gain
Setting range
–3.00 dB to +20.00 dB
Setting resolution
0.01 dB
Clipping level
upper and lower limit can be set
separately
0 % to 100 %
Maximum output voltage
see “Output voltage “ in section “Differential analog I/Q outputs”
7
The magnitude of the sum of output voltage and bias voltage must not exceed 4 V.
8
“Optimize internal I/Q impairments for RF output” switched off.