beautypg.com

Envelope tracking (r&s, Smw-k540 option) – Atec RohdeSchwarz-SMW200A User Manual

Page 22

background image

Version 01.02, September 2013

22

Rohde & Schwarz R&S

®

SMW200A Vector Signal Generator

Bias voltage (single-ended and differential) EMF

–3.6 V to +3.6 V

7

Resolution

2

mV

Uncertainty

1 % + 4 mV

Offset voltage

Differential

EMF

–300 mV to +300 mV

Resolution

0.1

mV

Uncertainty

1 % + 0.1 % × bias voltage + 1 mV

Differential signal balance

at R

L

= 50 Ω, output voltage > 0.5 V (V

p

)

Magnitude

up to 10 MHz

< 0.2 dB, 0.05 dB (meas.)

up to 80 MHz

0.2 dB (meas.)

Frequency response

8

at

R

L

= 50 Ω, output voltage > 0.5 V (V

p

)

Magnitude

up to 10 MHz

0.02 dB (meas.)

up to 80 MHz

0.03 dB (meas.)

Envelope tracking (R&S

®

SMW-K540 option)

With this option, the analog I/Q outputs can be used to generate an analog signal corresponding to the envelope of the I/Q signal to
test envelope tracking modulators..

This option can be installed once if the instrument is equipped with the R&S

®

SMW-B13 option. If the instrument is equipped with the

R&S

®

SMW-B13T option, envelope tracking can be used either on signal path A or B with one R&S

®

SMW-K540 option. For envelope

tracking to be used on signal paths A and B simultaneously, two R&S

®

SMW-K540 must be installed.

For each R&S

®

SMW-K540 option to be installed, a R&S

®

SMW-K16 option must be installed, and the instrument must be equipped

with at least one baseband generator (R&S

®

SMW-B10 option)

General
Envelope voltage adaption

auto, manual

Output type

single-ended, differential

Bias voltage

see section “Differential analog I/Q outputs”

Offset voltage

see section “Differential analog I/Q outputs”

Envelope to RF delay

Setting range

–500 ns to +500 ns

Setting resolution

1 ps

Shaping

off, linear, from table, polynomial,
dethroughing

Envelope voltage adaption mode auto
Power Amplifier Input Power P

in

Setting range

–145.00 dB to +30.00 dB

Setting resolution

0.01 dB

Power Amplifier Supply Voltage V

cc

see “Output voltage “ in section “Differential analog I/Q outputs”

DC Modulator gain

0.00 dB to +10.00 dB

Envelope voltage adaption mode manual
Pre-gain

Setting range

–20.00 dB to 0.00 dB

Setting resolution

0.01 dB

Post-gain

Setting range

–3.00 dB to +20.00 dB

Setting resolution

0.01 dB

Clipping level

upper and lower limit can be set
separately

0 % to 100 %

Maximum output voltage

see “Output voltage “ in section “Differential analog I/Q outputs”

7

The magnitude of the sum of output voltage and bias voltage must not exceed 4 V.

8

“Optimize internal I/Q impairments for RF output” switched off.