Dynamic Range Specifications

1 dB gain compression (two-tone)

Total power at mixer input

RF/MW

(Option 503, 507, 513, 526)

20 MHz to 26.5 GHz

+9 dBm nominal

Total power at mixer input

Millimeter-wave

(Option 532, 544)

20 MHz to 26.5 GHz

+6 dBm nominal

26.5 to 44 GHz

0 dBm nominal

Total power at preamp input

Preamp on

10 MHz to 3.6 GHz
3.6 to 26.5 GHz

Tone spacing: 100 kHz to 20 MHz
Tone spacing: > 70 MHz

> 26.5 GHz

–14 dBm nominal

–28 dBm nominal
–20 dBm nominal
–30 dBm nominal

Displayed average noise level (DANL)

(Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C)

Specification

Typical

RF/MW

(Option 503, 507, 513, 526)

1 to 10 MHz

–147 dBm

–149 dBm

10 MHz to 2.1 GHz

–148 dBm

–150 dBm

2.1 to 3.6 GHz

–147 dBm

–149 dBm

3.6 to 7.0 GHz

–147 dBm

–149 dBm

7.0 to 13.6 GHz

–143 dBm

–147 dBm

13.6 to 17.1 GHz

–137 dBm

–142 dBm

17.1 to 22 GHz

–137 dBm

–142 dBm

22 to 26.5 GHz

–134 dBm

–140 dBm

Preamp on, RF/MW

(Option 503, 507, 513, 526)

10 MHz to 2.1 GHz

–161 dBm

–163 dBm

2.1 to 3.6 GHz

–160 dBm

–162 dBm

3.6 to 7.0 GHz

–160 dBm

–162 dBm

7.0 to 13.6 GHz

– 160 dBm

–163 dBm

13.5 to 17.1 GHz

–157 dBm

–155 dBm

17.0 to 20.0 GHz

–155 dBm

–159 dBm

20.0 to 26.5 GHz

–150 dBm

–156 dBm

Millimeter-wave

(Option 532, 544)

9 kHz to 1 MHz

–130 dBm

1 MHz to 1.2 GHz

–152 dBm

–155 dBm

1.2 to 2.1 GHz

–151 dBm

–154 dBm

2.1 to 3.6 GHz

–149 dBm

–152 dBm

3.5 to 4.2 GHz

–144 dBm

–147 dBm

4.2 to 8.4 GHz

–145 dBm

–150 dBm

8.3 to 13.6 GHz

–147 dBm

–150 dBm

13.5 to 20 GHz

–145 dBm

–148 dBm

20 to 26.5 GHz

–142 dBm

–145 dBm

26.4 to 34 GHz

–140 dBm

–144 dBm

34.4 to 44 GHz

–135 dBm

–140 dBm

1. N is the LO multiplication factor.