beautypg.com

Atec Agilent-N9010A User Manual

Page 14

background image

14

Figure 4. Nominal phase noise at different center frequencies for millimeter-wave EXA (Option 532 or 544)

Phase noise

1

Offset

Specification

Typical

Millimeter-wave

100 Hz

–84 dBc/Hz

–88 dBc/Hz

(Option 532, 544)

1 kHz

–101 dBc/Hz nominal

Noise sidebands

10 kHz

–103 dBc/Hz

–106 dBc/Hz

(20 to 30 °C, CF = 1 GHz)

100 kHz

–115 dBc/Hz

–116 dBc/Hz

1 MHz

–135 dBc/Hz

–137 dBc/Hz

10 MHz

–149 dBc/Hz nominal

1. For nominal phase noise values with the millimeter-wave EXA (Option 532 or 544), refer to Figure 4.

Nominal phase noise at different center frequencies

with RBW selectivity curves, optimized phase noise versus offset frequency

SSB phase noise (dBc/Hz)

Freq (kHz)

Option MPB, microwave preselector bypass

1

Frequency range
N9010A-507

3.6 to 7 GHz

N9010A-513

3.6 to 13.6 GHz

N9010A-526

3.6 to 26.5 GHz

N9010A-532

3.6 to 32 GHz

N9010A-544

3.6 to 44 GHz

1. When Option MPB is installed and enabled, some aspects of the analyzer performance changes. Please refer to the EXA specification guide for more

details.