beautypg.com

Atec Agilent-N9010A User Manual

Page 8

background image

8

Input attenuation switching uncertainty

Specifications

Additional information

Attenuation > 2 dB, preamp off

50 MHz (reference frequency)

± 0.20 dB

± 0.08 dB typical

Relative to 10 dB (reference

setting)

9 kHz to 3.6 GHz

± 0.3 dB nominal

3.5 to 7.0 GHz

± 0.5 dB nominal

6.9 to 13.6 GHz

± 0.7 dB nominal

13.5 to 26.5 GHz

± 0.7 dB nominal

> 26.5 GHz

± 1.0 dB nominal

Total absolute amplitude accuracy

(10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp

Time = Accy, any reference level, any scale, σ = nominal standard deviation)

At 50 MHz

At all frequencies

9 kHz to 3.6 GHz

± 0.40 dB

± (0.40 dB + frequency response)

± 0.27 dB (95th percentile ≈ 2σ)

Preamp on

100 kHz to 3.6 GHz

± (0.39 dB + frequency response)

Input voltage standing wave ratio (VSWR) (≥ 10 dB input attenuation)

Options 503,

507, 513, 526

Options 532, 544

10 MHz to 3.6 GHz

< 1.2:1 nominal

1.2:1 nominal

3.6 to 26.5 GHz

< 1.8:1 nominal

1.5:1 nominal

26.5 to 44 GHz

N/A

< 1.8:1 nominal

Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW)

1 Hz to 3 MHz RBW

± 0.10 dB

4, 5, 6, 8 MHz RBW

± 1.0 dB

Reference level

Range

Log scale

–170 to +23 dBm in 0.01 dB steps

Linear scale

Same as Log (707 pV to 3.16 V)

Accuracy

0 dB

Display scale switching uncertainty

Switching between linear and log

0 dB

Log scale/div switching

0 dB

Display scale fidelity

Between –10 dBm and –80 dBm

input mixer level

± 0.15 dB total

Trace detectors

Normal, peak, sample, negative peak, log power average, RMS average, and voltage average

Preamplifier (Option P03, P07, P13, P26, P32, P44)

Frequency range

Option P03

100 kHz to 3.6 GHz

Option P07

100 kHz to 7 GHz

Option P13

100 kHz to 13.6 GHz

Option P26

100 kHz to 26.5 GHz

Option P32

100 kHz to 32 GHz

Option P44

100 kHz to 44 GHz

Gain

100 kHz to 3.6 GHz

+20 dB nominal

3.6 to 7.0 GHz

+35 dB nominal

> 7 GHz

+40 dB nominal

Noise figure

100 kHz to 3.6 GHz

8 to 12 dB nominal (proportional to frequency)

3.6 to 8.4 GHz

9 dB nominal

8.4 to 13.6 GHz

10 dB nominal

> 13.6 GHz

DANL + 176.24 dB nominal