General specifications – Atec Agilent-N9010A User Manual
Page 16
16
General Specifications
Temperature range
Operating
0 to 55 °C
Storage
–40 to 70 °C
EMC
Complies with European EMC Directive 2004/108/EC
• IEC/EN 61326-1 or IEC/EN 61326-2-1
• CISPR Pub 11 Group 1, class A
• AS/NZS CISPR 11:2002
• ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Safety
Complies with European Low Voltage Directive 2006/95/EC
• IEC/EN 61010-1 3rd Edition
• Canada: CSA C22.2 No. 61010-1-12
• U.S.A.: UL 61010-1 3rd Edition
Acoustic statement (European Machinery Directive 2002/42/EC, 1.7.4.2u)
Acoustic noise emission
LpA < 70 dB
Operator position
Normal position
Per ISO 7779
Environmental stress
Samples of this product have been type tested in accordance with the Agilent Environmental Test Manual and verified to be
robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to,
temperature, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are
similar to MILPRF-28800F Class 3.
Power requirements
Voltage and frequency
100 to 120 V, 50/60/400 Hz
220 to 240 V, 50/60 Hz
Power consumption
On
Standby
350 W maximum
20 W
Display
Resolution
1024 x 768, XGA
Size
213 mm (8.4 in.) diagonal (nominal)
Data storage
Internal
≥ 80 GB nominal (removable solid-state drive)
External
Supports USB 2.0 compatible memory devices
Weight (without options)
Net
16 kg (35 lbs) nominal
Shipping
28 kg (62 lbs) nominal
Dimensions
Height
177 mm (7.0 in)
Width
426 mm (16.8 in)
Length
368 mm (14.5 in)
Warranty
The EXA signal analyzer is supplied with a standard 3-year warranty
Calibration cycle
The recommended calibration cycle is two years; calibration services are available through Agilent service centers