Amplitude Accuracy and Range Specifications

Amplitude range

Measurement range

Displayed average noise level (DANL) to +23 dBm

Input attenuator range

(10 Hz to 44 GHz)

Standard

Option FSA

0 to 60 dB in 10 dB steps
0 to 60 dB in 2 dB steps

Electronic attenuator (Option EA3)

Frequency range

10 Hz to 3.6 GHz

Attenuation range

Electronic attenuator range

Full attenuation range

(mechanical + electronic)

0 to 24 dB, 1 dB steps
0 to 84 dB, 1 dB steps

Maximum safe input level

Average total power

(with and without preamp)

+30 dBm (1 W)

Peak pulse power

< 10 μs pulse width, < 1 % duty cycle +50 dBm (100 W) and input attenuation ≥ 30 dB

DC volts

DC coupled

AC coupled

± 0.2 Vdc
± 100 Vdc

Display range

Log scale

0.1 to 1 dB/division in 0.1 dB steps
1 to 20 dB/division in 1 dB steps (10 display divisions)

Linear scale

10 divisions

Scale units

dBm, dBmV, dBμV, dBmA, dBμA, V, W, A