Frequency response

Specification

95th percentile (≈ 2σ)

(10 dB input attenuation, 20 to 30 °C, preselector centering applied, σ = nominal standard deviation)
RF/MW

(Option 503, 507, 513, 526)

9 kHz to 10 MHz

± 0.8 dB

± 0.4 dB

10 MHz

1

to 3.6 GHz

± 0.6 dB

± 0.21 dB

3.5 to 7.0 GHz

± 2.0 dB

± 0.69 dB

6.9 to 13.6 GHz

± 2.5 dB

13.5 to 22.0 GHz

± 3.0 dB

22.0 to 26.5 GHz

± 3.2 dB

Millimeter-wave (Option 532, 544)

9 kHz to 10 MHz

± 0.6 dB

± 0.28 dB

10 to 50 MHz

± 0.45 dB

± 0.21 dB

50 MHz to 3.6 GHz

± 0.45 dB

± 0.20 dB

3.5 to 5.2 GHz

± 1.7 dB

± 0.91 dB

5.2 to 8.4 GHz

± 1.5 dB

± 0.61 dB

8.3 to 13.6 GHz

± 2.0 dB

± 0.61 dB

13.5 to 17.1 GHz

± 2.0 dB

± 0.67 dB

17.0 to 22.0 GHz

± 2.0 dB

± 0.78 dB

22.0 to 26.5 GHz

± 2.5 dB

± 0.72 dB

26.4 to 34.5 GHz

± 2.5 dB

± 1.11 dB

34.4 to 44 GHz

± 3.2 dB

± 1.42 dB

Preamp on (P03, P07, P13, P26)
RF/MW

(Option 503, 507, 513, 526)

100 kHz to 3.6 GHz

± 0.28 dB nominal

3.6 to 7.0 GHz

± 0.67 dB nominal

7.0 to 26.5 GHz

± 0.80 dB nominal

Preamp on (P03, P07, P32, P44)
Millimeter-wave (Option 532, 544)

100 kHz to 3.6 GHz

± 0.28 dB nominal

3.5 to 8.4 GHz

± 0.67 dB nominal

8.4 to 26.5 GHz

± 0.80 dB nominal

26.4 to 44 GHz

± 0.80 dB nominal

1. DC coupling required to meet specifications below 50 MHz. With AC coupling, specifications apply at frequencies of 50 MHz and higher. Statistical

observations at 10 MHz with AC coupling show that most instruments meet the DC-coupled specifications, however, a small percentage of

instruments are expected to have errors exceeding 0.5 dB at 10 MHz at the temperature extreme. The effect at 20 to 50 MHz is negligible but not

warranted.