Atec Agilent-N9010A User Manual
Page 20
20
Option B25 (standard) 25 MHz analysis bandwidth
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)
Center frequency (GHz)
Span (MHz)
Preselector
Max. error
RMS
≤ 3.6
10 to ≤ 25
n/a
± 0.45 dB
0.051 dB nominal
> 3.6
10 to ≤ 25
on
0.45 dB nominal
> 3.6
10 to ≤ 25
off
1
± 0.45 dB
0.071 dB nominal
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
Span (MHz)
Preselector
Peak-to-peak
RMS
0.02 ≤ f < 3.6
≤ 25
n/a
0.6°
0.14°
≥ 3.6
≤ 25
off
1
1.9°
0.4 °
≥ 3.6 (Option ≤ 526)
≤ 25
on
4.5°
1.2°
Data acquisition (25 MHz IF path)
Time record length (IQ pairs) IQ
Analyzer
4,000,000 IQ sample pairs
89600 software or N9064A
32-bit packing
64-bit packing
Memory
Option DP2, B40 or MPB
536 MSa
268 MSa
2 GB
None of the above
4,000,000 IQ sample pairs (independent of data packing)
Sample rate at ADC
Option DP2, B40 or MPB
100 MSa/s
None of the above
90 MSa/s
ADC resolution
Option DP2, B40 or MPB
16 bits
None of the above
14 bits
Option B40 40 MHz analysis bandwidth
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C), nominal
Center frequency (GHz)
Span (MHz)
Preselector
Max. error
RMS
0.03 ≤ f < 3.6
≤ 40
n/a
± 0.3 dB
0.08 dB
3.6 ≤ f ≤ 26.5
≤ 40
off
1
± 0.25 dB
0.08 dB
> 26.5
≤ 40
off
1
± 0.25 dB
0.12 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
Span (MHz)
Preselector
Peak-to-peak
RMS
0.02 ≤ f < 3.6
40
n/a
0.2°
0.05°
≥ 3.6
40
off
1
5°
1.4°
Data acquisition (40 MHz IF path)
Time record length (IQ pairs) IQ analyzer 4,000,000 samples (I/Q pairs)
89600 VSA software or N9064A VXA
32-bit packing
64-bit packing
2 GB total memory (nominal)
Length (IQ sample pairs)
536 MSa
268 MSa
Length (time units)
Samples/(span x 1.28) (nominal)
Sample rate
At ADC
200 Msa/s
IQ pairs
Span x 1.28 (nominal)
ADC resolution
12 bits
1. Option MPB is installed and enabled.