Frequency and Time Specifications
Frequency range
DC coupled
AC coupled
Option 503
10 Hz to 3.6 GHz
10 MHz to 3.6 GHz
Option 507
10 Hz to 7 GHz
10 MHz to 7 GHz
Option 513
10 Hz to 13.6 GHz
10 MHz to 13.6 GHz
Option 526
10 Hz to 26.5 GHz
10 MHz to 26.5 GHz
Option 532
10 Hz to 32 GHz
NA
Option 544
10 Hz to 44 GHz
NA
Band
LO multiple (N)
0
1
10 Hz to 3.6 GHz
1
1
3.5 to 7.0 GHz
1
1
3.5 to 8.4 GHz
2
2
8.4 to 13.6 GHz
3
2
13.5 to 17.1 GHz
4
4
17 to 26.5 GHz
5
4
26.4 to 34.5 GHz
6
8
34.4 to 44 GHz
Frequency reference
Accuracy
± [(time since last adjustment x aging rate) + temperature stability + calibration
accuracy]
Aging rate
Option PFR
Standard
± 1 x 10
–7
/ year
± 1 x 10
–6
/ year
± 1.5 x 10
–7
/ 2 years
Temperature stability
20 to 30 °C
Full temperature range
Option PFR
Standard
± 1.5 x 10
–8
± 2 x 10
–6
± 5 x 10
–8
± 2 x 10
–6
Achievable initial calibration accuracy
Option PFR
Standard
± 4 x 10
–8
± 1.4 x 10
–6
Example frequency reference accuracy
(with Option PFR)
= ± (1 x 10
–7
+ 5 x 10
–8
+ 4 x 10
–8
)
1 year after last adjustment
= ± 1.9 x 10
–7
Residual FM
Option PFR
Standard
≤ (0.25 Hz x N) p-p in 20 ms nominal
≤ (10 Hz x N) p-p in 20 ms nominal
See band table above for N (LO Multiple)
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW
+ 2 Hz + 0.5 x horizontal resolution
1
)
Marker frequency counter
Accuracy
± (marker frequency x frequency reference accuracy + 0.100 Hz)
Delta counter accuracy
± (delta frequency x frequency reference accuracy + 0.141 Hz)
Counter resolution
0.001 Hz
Frequency span (FFT and swept mode)
Range
0 Hz (zero span), 10 Hz to maximum frequency of instrument
Resolution
2 Hz
Accuracy
Swept
FFT
± (0.25 % x span + horizontal resolution)
± (0.10 % x span + horizontal resolution)
1. Horizontal resolution is span/(sweep points – 1).