Magnitude, Phase – Atec Agilent-N5230A User Manual
Page 8
8
N5230A Option 140/145/146/240/245/246
Corrected system performance with 3.5 mm connectors
1
(continued)
Table 4. N4433A Electronic calibration module
N5230A
• Option 140 or 240 standard test set and standard power range
•
Option
145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second
source
Applies to the N5230A Option 140/145/146/240/245/246 analyzers, N4433A
electronic calibration module, 85131F flexible test port cable set, and a full 4-port calibration.
Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature.
Specification
(dB)
Description
10 to
500 MHz to
2 to
8 to
13.5 to
500 MHz
2 GHz
8 GHz
13.5 GHz
20 GHz
Directivity 52 52 47 45 45
Source match
42
42
39
31
31
Load match
45
45
41
35
35
Reflection
±0.060
±0.060
±0.090
±0.040
±0.180
tracking
(+0.01/°C)
(+0.01/°C)
(+0.02/°C)
(+0.03/°C)
(+0.03/°C)
Transmission
±0.045
±0.039
±0.055
±0.127
±0.160
tracking
(+0.01/°C)
(+0.01/°C)
(+0.02/°C)
(+0.03/°C)
(+0.03/°C)
Transmission uncertainty (specifications)
1. From 300 kHz to 10 MHz, performance is characterized as “typical”. To generate these typical values,
please download our free Uncertainty Calculator from
www.agilent.com/find/na_calculator.
2. All of the curves are for 2-port calibrations. Multiport uncertainties are currently not supported.
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
N5230A Option 140/145/146/240/245/246 full 2-port
2
cal using N4433A
S11 = S22 = 0
Source power = -5 dBm
Uncertainty (dB)
■
10 to 500 MHz
♦
500 MHz to 2 GHz
▲
2 to 8 GHz
●
8 to 20 GHz
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
N5230A Option 140/145/146/240/245/246 full 2-port
2
cal using N4433A
S11 = S22 = 0
Source power = -5 dBm
Uncertainty (degrees)
■
10 to 500 MHz
♦
500 MHz to 2 GHz
▲
2 to 8 GHz
●
8 to 20 GHz