beautypg.com

Phase, Magnitude – Atec Agilent-N5230A User Manual

Page 7

background image

7

Table 3. 85052B Calibration kit (continued)

N5230A

• Option 140 or 240 standard test set and standard power range

Option

145 or 245 configurable test set and extended power range

• Option 146 or 246 configurable test set and extended power range with internal

second

source

Reflection uncertainty (specifications)

Phase

0.1

1

10

100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

N5230A Option 140/145/146/240/245/246 with 85052B

S11 = S22 = 0
Source power = -5 dBm

Uncertainty (degrees)

10 to 500 MHz

500 MHz to 2 GHz

2 to 8 GHz

8 to 20 GHz

Magnitude

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

N5230A Option 140/145/146/240/245/246 with 85052B

S21 = S12 = 0
Source power = -5 dBm

Uncertainty (linear)

10 to 500 MHz

500 MHz to 2 GHz

2 to 8 GHz

8 to 20 GHz

Phase

0

2

4

6

8

10

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

N5230A Option 140/145/146/240/245/246 with 85052B

S21 = S12 = 0
Source power = -5 dBm

Uncertainty (deg)

10 to 500 MHz

500 MHz to 2 GHz

2 to 8 GHz

8 to 20 GHz