Atec Agilent-N5230A User Manual
Page 28
28
Table 16. Cycle time vs number of points
Applies to the preset condition (correction off) except for the following changes:
• CF = 10 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)
IF Bandwidth (Hz)
Number of points
Cycle time (ms)
1
30,000 3
2.88
11
3.50
51
3.91
101
5.29
201
8.75
401
15.66
801
29.46
1,601
57.73
6,401
221.04
16,001
549.71
100,000 3
2.87
11
2.82
51
2.86
101
2.96
201
4.02
401
6.23
801
10.65
1,601
19.49
6,401
70.96
16,001
173.78
600,000 3
2.84
11
2.84
51
2.87
101
3.03
201
3.14
401
3.51
801
4.22
1,601
6.22
6,401
19.35
16,001
45.12
1. Cycle time includes sweep and retrace time.
Table 17. Data transfer time (ms)
Number
of
points
201
401
1601
16,001
SCPI over GPIB (program executed on external PC)
32-bit floating point
6
10
33
313
64-bit floating point
10
18
65
622
ASCII
53
105
406
4032
CPI over SIU/LAN or TCP/IP socket (program executed in the analyzer)
32-bit floating point
1
2
2.5
7
64-bit floating point
2
2
3
10
ASCII
11
20
73
720
COM (program executed in the analyzer)
32-bit floating point
< 0.2
0.2
0.3
0.9
Variant
type
0.6 1 3.2 32
DCOM over LAN (program executed on external PC)
32-bit floating point
< 0.8
1
1.6
7.5
Variant
type
1.9 3 8.9 82
Note: Specifications for recall and sweep speed are not provided for the N5230A analyzers.