Measurement throughput summary – Atec Agilent-N5230A User Manual
Page 26
26
Measurement Throughput Summary
Table 14. Typical cycle time
1
(ms) for measurement completion
Number of Points
201 401 801 1601
16,001
Start 8 GHz, stop 18 GHz, 600 kHz IF bandwidth
Uncorrected
21.148 21.743 23.01 25.198 54.836
4-Port cal
74.597
82.296
296.5
307.75 538.646
Start 300 kHz, stop 10 GHz, 600 kHz IF bandwidth
Uncorrected 19.814
22.801
24.973
29.01
67.733
4-Port
cal
69.752 85.111 100.125 129.347 480.711
Start 300 kHz, stop 20 GHz, 600 kHz IF bandwidth
Uncorrected 32.575
34.7
39.237
43.155
69.625
4-Port
cal
121.254 133.626 157.506 179.223 487.779
Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected
38.083 51.816 55.488 56.36 184.154
4-Port
cal
143.271 201.814 215.056 230.133 934.161
Start 300 kHz, stop 10 GHz, 100 kHz IF bandwidth
Uncorrected
37.03 42.532 45.122 46.729 198.683
4-Port
cal
137.431 162.37 194.13 192.182 906.768
Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected
44.98 69.408 87.161 92.475 198.792
4-Port
cal
169.041 268.877 343.898 369.526 914.963
Start 8 GHz, stop 18 GHz, 50 kHz IF bandwidth
Uncorrected 42.171
70.09
88.702
90.981
371.611
4-Port
cal
157.107 271.791 351.517 368.02 1532.609
Start 300 kHz, stop 10 GHz, 50 kHz IF bandwidth
Uncorrected 43.713
61.41
66.878
69.373
385.04
4-Port cal
163.58 238.267 259.687 279.816 1580.761
Start 300 kHz, stop 20 GHz, 50 kHz IF bandwidth
Uncorrected
48.673 80.798 124.605 147.303 388.46
4-Port
cal
184.429 313.392 493.142 587.548 1587.839
1. Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with
DISPLAY:ENABLE OFF. Add 21 ms for display on. Data for one trace (S11) measurement.