N5230A Option 140/145/146/240/245/246
Corrected system performance with 3.5 mm connectors

1

Note: For any S

ii

reflection measurement:

S

jj

= 0

For any S

ij

transmission measurement:

S

ji =

S

ij

when

S

ij

≤ 1

S

ji =

1/S

ij

when

S

ij

≥ 1

S

kk

=

0

for all k

Table 3. 85052B Calibration kit

N5230A

• Option 140 or 240 standard test set and standard power range

Option

145 or 245 configurable test set and extended power range

• Option 146 or 246 configurable test set and extended power range with internal

second

source

Applies to the N5230A Option 140/145/146/240/245/246 analyzers, 85052B (3.5mm)
calibration kit, 85131F flexible test port cable set, and a full 4-port calibration. Also applies
to the following condition: Environmental temperature 23° ±3 °C, with < 1 °C deviation
from calibration temperature.

Specification (dB)

Description

10 to

500 MHz to

2 to

8 to

13.5 to

500 MHz

2 GHz

8 GHz

13.5 GHz

20 GHz

Directivity 48

48 44 44 44

Source match

40

40

33

31

31

Load match

48

48

44

44

44

Reflection

±0.003

±0.003

±0.003

±0.006

±0.006

tracking

(+0.01/°C)

(+0.01/°C)

(+0.02/°C)

(+0.03/°C)

(+0.03/°C)

Transmission

±0.017

±0.017

±0.062

±0.125

±0.125

tracking

(+0.01/°C)

(+0.01/°C)

(+0.02/°C)

(+0.03/°C)

(+0.03/°C)

Transmission uncertainty (specifications)

1. From 300 kHz to 10 MHz, performance is characterized as “typical”. To generate these typical values,
please download our free Uncertainty Calculator from

www.agilent.com/find/na_calculator.

Magnitude

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

N5230A Option 140/145/146/240/245/246 with 85052B

S11 = S22 = 0
Source power = -5 dBm

Uncertainty (dB)

10 to 500 MHz

500 MHz to 2 GHz

2 to 8 GHz

8 to 20 GHz