Magnitude – Atec Agilent-N5230A User Manual
Page 6
6
N5230A Option 140/145/146/240/245/246
Corrected system performance with 3.5 mm connectors
1
Note: For any S
ii
reflection measurement:
S
jj
= 0
For any S
ij
transmission measurement:
S
ji =
S
ij
when
S
ij
≤ 1
S
ji =
1/S
ij
when
S
ij
≥ 1
S
kk
=
0
for all k
Table 3. 85052B Calibration kit
N5230A
• Option 140 or 240 standard test set and standard power range
•
Option
145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second
source
Applies to the N5230A Option 140/145/146/240/245/246 analyzers, 85052B (3.5mm)
calibration kit, 85131F flexible test port cable set, and a full 4-port calibration. Also applies
to the following condition: Environmental temperature 23° ±3 °C, with < 1 °C deviation
from calibration temperature.
Specification (dB)
Description
10 to
500 MHz to
2 to
8 to
13.5 to
500 MHz
2 GHz
8 GHz
13.5 GHz
20 GHz
Directivity 48
48 44 44 44
Source match
40
40
33
31
31
Load match
48
48
44
44
44
Reflection
±0.003
±0.003
±0.003
±0.006
±0.006
tracking
(+0.01/°C)
(+0.01/°C)
(+0.02/°C)
(+0.03/°C)
(+0.03/°C)
Transmission
±0.017
±0.017
±0.062
±0.125
±0.125
tracking
(+0.01/°C)
(+0.01/°C)
(+0.02/°C)
(+0.03/°C)
(+0.03/°C)
Transmission uncertainty (specifications)
1. From 300 kHz to 10 MHz, performance is characterized as “typical”. To generate these typical values,
please download our free Uncertainty Calculator from
www.agilent.com/find/na_calculator.
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
N5230A Option 140/145/146/240/245/246 with 85052B
S11 = S22 = 0
Source power = -5 dBm
Uncertainty (dB)
■
10 to 500 MHz
♦
500 MHz to 2 GHz
▲
2 to 8 GHz
●
8 to 20 GHz