Table 5. Uncorrected system performance

1

Directivity

Specifications

Typicals

Options 140, 145, 146,

Options 140, 145, 146,

240, 245, 246

240, 245,246

300 kHz to 10 MHz

–23 dB

10 MHz to 1 GHz

–28 dB

1 to 3 GHz

–25 dB

3 to 5 GHz

–20 dB

5 to 11.5 GHz

–17 dB

11.5 to 13.5 GHz

–15 dB

13.5 to 20 GHz

–15 dB

Source match
300 kHz to 10 MHz

–8 dB

10 MHz to 1 GHz

–12 dB

1 to 3 GHz

–12 dB

3 to 5 GHz

–12 dB

5 to 10.5 GHz

–12 dB

10.5 to 11.5 GHz

–10 dB

11.5 to 13.5 GHz

–8 dB

13.5 to 20 GHz

–8 dB

Load match
300 kHz to 10 MHz

–9 dB

10 MHz to 1 GHz

–20 dB

1 to 3 GHz

–20 dB

3 to 5 GHz

–18 dB

5 to 11.5 GHz

–12 dB

11.5 to 13.5 GHz

–7 dB

13.5 to 16 GHz

–7 dB

16 to 20 GHz

–7.5 dB

Crosstalk

2

300 kHz to 5 MHz

–70 dB

5 to 10 MHz

–100 dB

10 to 45 MHz

–110 dB

45 MHz to 4 GHz

–122 dB

4 to 6 GHz

–123 dB

6 to 10.5 GHz

–120 dB

10.5 to 13.5 GHz

–115 dB

13.5 to 15 GHz

–115 dB

15 to 20 GHz

–110 dB

1. Specifications apply over environmental temperature of 25 °C ±5 °C with less than 1 °C variation
from calibration temperature.
2. Measurement conditions: normalized to a thru, measured with two shorts, 10 Hz IF bandwidth,
averaging factor of 8, alternate mode source power set to the lesser of the maximum power out
or the maximum receiver power.