Table 13. Analyzer environment and dimensions

Description Supplemental

information

General environmental
EMC

Complies with European EMC directive 89/336/EEC, amended by 93/68/

EEC

• IEC/EN 61326

• CISPR Pub 11 Group 1, class A

• AS/NZS CISPR11:2002

ICES/NMB-001

Safety

Complies with European Low Voltage Directive 73/23/EEC, amended by

93/68/EEC

• IEC/EN 61010-1:2001

• Canada: CSA C22.2 No. 61010-1:2001

• USA: UL 61010-1

Operating environment

Temperature

0 to +40 °C

Instrument powers up and displays no error messages within this

temperature range (except for "source unleveled" error message that

may occur at temperatures outside the specified performance

temperature range of 25 ± 5 °C).

Error-corrected temperature range

23 °C ± 3 °C with less than 1 °C deviation from calibration temp.

Relative humidity

Type tested 0 to 95% at 40 °C, non-condensing

Altitude

0 to 4600 m (15,000 ft.)

Non-operating storage environment

Temperature

–40 to +70 °C

Cabinet dimensions

Height

Width

Depth

Excluding front and rear panel hardware and feet

267 mm 426 mm 427 mm

10.5 in

16.75 in 16.8 in

As shipped - includes front panel connectors,

280 mm 435 mm 470 mm

rear panel bumpers, and feet.

11 in

17.10 in 18.5 in

As shipped plus handles

280 mm 458 mm 501 mm

11 in

18 in

19.7 in

As shipped plus rack-mount flanges

280 mm 483 mm 470 mm

11 in

19 in

18.5 in

As shipped plus handles and rack-mount flanges

280 mm 483 mm 501 mm

11 in

19 in

19.7 in

Weight
Net

N5230A

24.9 kg (55 lb), nominal

Shipping

N5230A

36.3 kg (80 lb), nominal